Specifications
Chapter 1 Introduction
MiniCapt Remote Microbial Air Sampler Operations Manual
Page 1-5
Specifications
Table 1-1
MiniCapt Remote Microbial Air Sampler Specifications
25R
50R
100R
Sampling flow rate
25 Liters/minute
50 Liters/minute
100 Liters/minute
Communication
Modbus TCP
Dimensions (h, w, d)
Enclosure: 7.2 x 5.9 x 6.3 in (18 x 15 x 15 cm)
Weight
7.4 lb (3.4 kg) without BioCapt Microbial Impactor
Operating environment
Temperature: 32 - 104 °F (0 - 40 °C); Humidity: 10 – 90% RH
Maximum altitude
6,562 ft (2000 m)
Calibration interval
Recommended every 12 months
Utility Requirements
Electrical rating
100 – 240 V, 50 – 60 Hz, 1.0 A
Voltage fluctuation
AC input voltage fluctuation shall not 10%
Installation Requirements
Indoor use only
Pollution degree 2
Over-voltage Category II
Ordinary Protection (Not protected against harmful ingress of
moisture.)
Class I Equipment (Electrical earth ground from the mains power
source to the product input is required for safety.)
BioCapt Impactor Head
Materials
AISI 316L Stainless Steel
Number of inlets
20 (precision-cut slits)
Accessories
Compressed Gases Sampling Kit
Isolator Monitoring Kit
Remote ISP Sampling Kit