14

3.0 Front Panel: Data Entry and Display 

 

 

The front panel consists of a user interface, which includes an alphanumeric display and a 
membrane keypad to select operational modes and alarm settings. The display is a two line, 20 
character each, liquid crystal display (LCD). The display is backlit to allow easy viewing over a wide 
range of lighting conditions. 

The lower line on the LCD is used to signify the function of the “soft keys” marked “A”, “B” and “C”. 
The “soft key” current labels are displayed in the lower line of the LCD. If you press these keys, 
then the function displayed above it will be performed. 

The main keypad consists of eight “hard” keys whose function does not change. These keys are 
used for basic control and programming of the LabTec.  

The A B and C keys along with the Exit button are the only ones used in normal operation.  
Pressing any key will take one out of Serial Mode, the mode used with this unit. 

If the unit displaying something other than 

“Serial CW 160 ml/m”

 on the top line, press the Exit 

button to reach 

“Mode Select Serial”,

 and then press C to select, and A to execute to return to the 

proper screen. 

Two LED’s are also on the front panel, just to the left of the main keypad. These indicate the current 
pump status. A green light indicated the pump is in motion; the red light indicated that the pump has 
stopped. 

 

 

 

Содержание SciLog LabTec Series

Страница 1: ...LabTec Methadone Dispenser Installation Operation Maintenance Instructions Manual...

Страница 2: ...cy of product development and although the Company reserves the right to change specification it attempts to keep customers informed of any alterations This publication is for general information only...

Страница 3: ...TALLING SERVICING CLEANING OR REMOVING THE FUSE FAILURE TO DO SO COULD RESULT IN BODILY HARM AND OR PROPERTY DAMAGE CAUTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES WARNING ONL...

Страница 4: ...OU AVANT DE RETIRER LE FUSIBLE NE PAS OBSERVER CES PR CAUTIONS RISQUERAIT DE CAUSER DES BLESSURES CORPORELLES OU ET D ENDOMMAGER L QUIPEMENT PRUDENCE SOYEZ PRUDENT LORSQUE VOUS MANIPULEZ DES APPAREIL...

Страница 5: ...RT A LABTEC METHADONE HARDWARE 16 1 0 Hardware Overview 16 2 0 Pump Head Motor 16 3 0 Front Panel Data Entry and Display 16 4 0 Back Panel Interface Options 18 5 0 LabTec Cleaning Maintenance RH1 CKC...

Страница 6: ...fied by the manufacturer Maintenance Cleaning The LabTec Methadone requires routine daily maintenance The pump head in particular requires daily flushing with an approved cleaning solution and distill...

Страница 7: ...Birtley Co Durham Oxnard California USA 93030 DH3 2SF England toll free 877 784 2234 phone 44 0 191 4105121 phone 1 805 604 3400 fax 44 0 191 4105312 fax 1 805 604 3401 email bioscience emea parker c...

Страница 8: ...e cover S3 Male DB9 connector Not utilized on LabTec Methadone Do not remove the cover External I O Female DB37 connector Not utilized on LabTec Methadone V Female DB15 connector Not utilized on LabTe...

Страница 9: ...ing the LabTec Operating Manual into the PC choose Install the software automatically and click Next The following screen appears Choose Continue Anyway and the driver will finish loading allowing you...

Страница 10: ...on Browse my computer on the next screen and browse to the CD drive that contains the LabTec Manual and click on OK 3 As mentioned earlier you will get a security warning about windows verification f...

Страница 11: ...SB Serial Device automatically D Plug in and power up the LabTec E Remove the short piece of tubing installed in the pump for shipping and retain it for use at when sending the system in for maintenan...

Страница 12: ...efore continuing with your pumping application immediately replace the methadone container with your cleaning solution press the A button twice to return to the normal screen and try dispensing again...

Страница 13: ...and a RH1 CKC piston pump head Materials of construction the pump body is made of Kynar the piston and cylinders are made of ceramic Both Kynar and ceramic are inert materials and do not interact reac...

Страница 14: ...splayed above it will be performed The main keypad consists of eight hard keys whose function does not change These keys are used for basic control and programming of the LabTec The A B and C keys alo...

Страница 15: ...done Please do not remove the covers Foot Switch P N 080 059 Male DB37 Labeled External I O Not utilized in the LabTec Methadone unit SciPres Disposable Pressure Sensors 3 RJ11 jacks one for each pres...

Страница 16: ...cally implement the correct communications parameters Check that the proper communications parameters are also implemented in the scale being used 4 3 PRESSURE SENSOR PORTS The LabTec Methadone unit d...

Страница 17: ...lution reservoir until you are ready to prime the pump with Methadone the following day Do not leave Methadone in the pump head overnight Preventive maintenance is very valuable and ensures a long ope...

Страница 18: ...uated cylinder or a balance with a 0 1 gm resolution or better B Enter the Calibration section of your clinical software and instruct the pump to dispense 200 mg This is equivalent to a 20 ml volume O...

Страница 19: ...and when doing so cut both pieces a foot longer than needed This can be done more frequently if the clinic wishes to This allows you to remove the tubing from the pump once a month and cutting of the...

Страница 20: ...20 Part B LabTec Methadone Software 1 0 MAIN MENU...

Страница 21: ...nu is entered which provides access to the Exec and Edit functions In the Edit submenu the parameters for the application are selected SERIAL In the Serial mode the LabTec is connected to PC running c...

Страница 22: ...22 2 0 SERIAL MODE...

Страница 23: ...8 bits Deviation from these settings will cause communication issues generating errors on the computer Please do not change these values unless instructed by Parker Tech Support or that of your clinic...

Страница 24: ...to dispense 2 00 grams send W0200 X REPEAT begins another dispensing cycle for the current dispense volume or weight F FORWARD sets the current pump direction to clockwise CW will be displayed R REVER...

Страница 25: ...eing a stuck pump head 4 0 Manual Mode Summary In the Manual Mode the LabTec Methadone can be manually operated The pump speed can be set in terms of motor speed while the pump is running Just press t...

Страница 26: ...ccurred in the middle of the day and can t afford to stop dispensing 1 Press the A button twice to clear the error and return to regular screen 1 Send alternating Prime and Empty or Flush commands rep...

Страница 27: ...nia USA 93030 toll free 877 784 2234 phone 1 805 604 3400 fax 1 805 604 3401 email bioscience na parker com www parker com bioprocessing Parker Hannifin Corporation Bioscience Division N A 2340 Eastma...