Description of Unit 

The Digital MDM Mixer is shown in Figure 1. The device permits the operator to adjust the percent of oxygen through a 
range of 100% to 30% (and thereby control the ratio of oxygen to nitrous oxide). The operator can also control the total 
volume of mixed gas up to 9.9 liters per minute. 

Mixture Control 

The  DIGITAL  MDM  mixer  precisely  mixes  oxygen  and  nitrous  oxide  from  local  cylinders  or  pipeline  gas  systems,  and 
delivers the electronically controlled mixture to the bag tee outlet. 

Built-In Safety 

A series of alarms (see Safety Alarm System page 7) alert the user to any problem condition within the DIGITAL MDM mixer 
unit or in the gas supply. 

Displays and Indicators (Figure 1)

 

Flow Display:

 The LPM FLOW display (4) normally displays the total gas flow rate. It also displays the individual nitrous 

oxide or oxygen flow rates when either the N

2

O or O

2

 flow display controls (6) are pressed. The gas flow being displayed is 

indicated in the flow indicator (5) by O

2

, TOTAL or N

2

O. 

% OXYGEN Display:

 The % OXYGEN display (3) indicates the oxygen concentration set point in the total flow. 

Gas Flow Indicators:

 Gas flow is indicated by bar graph LEDs (1). A green bar graph represents O

2

 while blue represents 

the N

2

O flow. The number of lit bars gives a graphical indication of the flow level of a particular gas. 

 

Figure 1: 

Digital MDM Mixer Front Panel (figure shows all LEDs illuminated)

 

 

 

Содержание Matrx Digital MDM

Страница 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Страница 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Страница 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Страница 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Страница 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Страница 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Страница 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Страница 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Страница 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Страница 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Страница 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Страница 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Страница 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Отзывы: