Specifications 

 

Physical 

Dimensions: 20.32cm W x 13.97cm H x 25.4cm D 
Weight: 2.7kg. 
 

Fittings 

Fresh Gas Outlet:  15mm ID x 22mm OD 
Nitrous Oxide Inlet:  Male SIS  
Oxygen Inlet:  Male SIS  
 

Environmental 

Storage Temperature: -10°F to 120°F (-17°C to 48°C) 
(Allow to stabilize to room temperature before operating.) 
Operating Temperature: 72°F nominal (22°C) 
Humidity: Ambient, non-condensing 
 

Gas Supply 

Oxygen Inlet 
Pressure Requirements: 

 

 344.7 kPa to 379 kPa 

Flow Requirements: 

 

30 LPM minimum flow 

 

Nitrous Oxide Inlet 

Pressure Requirements: 344.7 Kpa to 379 kPa 
Flow Requirements:  10 LPM minimum flow 
 

Gas Delivery 

Oxygen Flush:10 to 30 LPM (20 LPM nominal @ 344.7 
kPa input) 
Minimum Nitrous Oxide Concentration: 0% 
Maximum Nitrous Oxide Concentration: 70% 
Minimum Oxygen Concentration: 30% 
Maximum Oxygen Concentration: 100% 
Flow Range @ 100% Oxygen: 1.0 to 9.9 LPM 
Flow 

Range @ Any % Mixture: up to 9.9 LPM† 

(†Low end flow limited by 1.0 LPM oxygen limit) 

 

Electrical (220

–240 V) 

Input Line Frequency:  50

–60Hz 

Circuit Protection: 

Two 2.5 amp time delay fuses 

 

(Matrx PN 63882537) 

 

 

Performance and Accuracy

 (Controlled Variable 344.7 kPa 

– 379 kPa –Sources) 

Oxygen Flowrate from 1.0 to 9.9 ±0.5 LPM 

Nitrous Oxide Flowrate from 1.0 to 7.0 LPM ±0.5 LPM 

Nitrous Oxide Flowrate from 0.1 to 1.0 LPM +0.5 / -0.9 LPM 

Percent O

2

: ±5 % points @ N

2

O > 1.0 LPM 

Percent O

2

: +20 / -10 % points @ N

2

O < 1.0 LPM 

Pure Oxygen Flush: 20 LPM Nominal at 344.7 kPa Input 

 

Check out our website 

www.porterinstrument.com

 for the following topics.

 

Flowmeter Mounting Options, Flowmeter Accessories, Product Warranty and Registration 

And Downloading User Manuals 

 

 
 

Supplied in Australia by: 

 

Air Liquide Healthcare Pty Ltd: Unit 5/476 Gardeners Road, Alexandria, NSW 2015 

Tel: (02) 9364 7474  Fax: (02) 9364 7476 

www.airliquidehealthcare.com.au 

 

 
 

Содержание Matrx Digital MDM

Страница 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Страница 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Страница 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Страница 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Страница 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Страница 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Страница 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Страница 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Страница 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Страница 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Страница 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Страница 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Страница 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Отзывы: