13
10545100-Australia Rev.
– 08/2018
Specifications
Physical
Dimensions: 20.32cm W x 13.97cm H x 25.4cm D
Weight: 2.7kg.
Fittings
Fresh Gas Outlet: 15mm ID x 22mm OD
Nitrous Oxide Inlet: Male SIS
Oxygen Inlet: Male SIS
Environmental
Storage Temperature: -10°F to 120°F (-17°C to 48°C)
(Allow to stabilize to room temperature before operating.)
Operating Temperature: 72°F nominal (22°C)
Humidity: Ambient, non-condensing
Gas Supply
Oxygen Inlet
Pressure Requirements:
344.7 kPa to 379 kPa
Flow Requirements:
30 LPM minimum flow
Nitrous Oxide Inlet
Pressure Requirements: 344.7 Kpa to 379 kPa
Flow Requirements: 10 LPM minimum flow
Gas Delivery
Oxygen Flush:10 to 30 LPM (20 LPM nominal @ 344.7
kPa input)
Minimum Nitrous Oxide Concentration: 0%
Maximum Nitrous Oxide Concentration: 70%
Minimum Oxygen Concentration: 30%
Maximum Oxygen Concentration: 100%
Flow Range @ 100% Oxygen: 1.0 to 9.9 LPM
Flow
Range @ Any % Mixture: up to 9.9 LPM†
(†Low end flow limited by 1.0 LPM oxygen limit)
Electrical (220
–240 V)
Input Line Frequency: 50
–60Hz
Circuit Protection:
Two 2.5 amp time delay fuses
(Matrx PN 63882537)
Performance and Accuracy
(Controlled Variable 344.7 kPa
– 379 kPa –Sources)
Oxygen Flowrate from 1.0 to 9.9 ±0.5 LPM
Nitrous Oxide Flowrate from 1.0 to 7.0 LPM ±0.5 LPM
Nitrous Oxide Flowrate from 0.1 to 1.0 LPM +0.5 / -0.9 LPM
Percent O
2
: ±5 % points @ N
2
O > 1.0 LPM
Percent O
2
: +20 / -10 % points @ N
2
O < 1.0 LPM
Pure Oxygen Flush: 20 LPM Nominal at 344.7 kPa Input
Check out our website
www.porterinstrument.com
for the following topics.
Flowmeter Mounting Options, Flowmeter Accessories, Product Warranty and Registration
And Downloading User Manuals
Supplied in Australia by:
Air Liquide Healthcare Pty Ltd: Unit 5/476 Gardeners Road, Alexandria, NSW 2015
Tel: (02) 9364 7474 Fax: (02) 9364 7476
www.airliquidehealthcare.com.au