10545100-Australia Rev. 

– 08/2018 

REQUIREMENTS AND FEATURES 

Bag Tee (Breathing Bag) Requirements 

A bag tee (with attached breathing bag or capped breathing bag port) containing an air inlet valve is required for use with 
this unit. The bag tee is factory mounted on the unit or provided separately for remote mounting. Matrx by Parker Scavenging 
Nasal  Masks  (or  other  suitable  breathing  circuits 

–  see  Breathing  Circuits  page  5)  provide  the  patient  connection.  An 

Anaesthetic  Breathing  Circuit  incorporating  Scavenging  with  face  mask/nasal  hood  may  also  provide  the  patient 
connections. Refer to: Breathing Circuits on page 5. 

Use of Scavenging 

A patient breathing circuit that incorporates scavenging and a one-way exhalation valve must be used with this device. For 
applications other than Dental use, a separate Scavenger Interface must be used with this system for connection between 
patient breathing circuit and facility evacuation system. If used in a dental application, the scavenging tube of the patient 
breathing circuit may connect directly to chair suction and a scavenging interface may not be required. 

A Scavenger interface contains an internal pathway for gas passage permitting waste gas to follow a preferential path to 
vacuum. 

Scavenger Interfaces must meet the following criteria:

 

 

Must be open to ambient air at either top or bottom of interface (via perforations) 

 

Must contain a sintered filter for particulate capture 

 

A barb tail connector for connection to wall evacuation system  

 

A  rigid  conical  19mm  connection  for  connection  to  the  breathing  circuit  scavenging  tube  (refer  AS/NZ  2496: 
Breathing attachments for anaesthetic purpose for human use) 

Note: Follow specific manufacturer Instructions for Use for individual Scavenger Interfaces 
Monitor for N

2

O in the patient treatment area to ensure that controls are effective in achieving low PPM (parts per million) 

exposure (TWA = 25ppm in Australia). Contact your dealer for details on monitors and testing. 
 

Room Ventilation 

Mechanical  ventilation  requirements 

shall  conform  to 

Australian  Standard 

AS1668.2:  The  use  of  ventilation  and  air-

conditioning  in  buildings Part 2: Mechanical  ventilation in buildings, Section 5: Enclosures used for particular healthcare 
functions  or  those  specified  in

 

The  Australasian  Healthcare  Facility  Guidelines  and/or  State  and  Territory  Engineering 

Services/Healthcare Infrastructure Guidelines and Specifications on air change requirements as applicable. 
 

Facility Piping 

The user MUST perform tests independent of the dealer or contractor to verify that all pipelines are connected correctly. If 
you are unfamiliar with piping systems, refer to Australian Standard AS2896: 2011 Medical gas systems 

– Installation and 

testing  on  non-flammable  medical  gas  pipeline  systems. The  user  has  the  ultimate  responsibility  for  properly  connected 
pipelines essential to patient safety. 
 

Sleeve Index System (SIS) 

Gas connections in the DIGITAL MDM mixer are mechanically indexed using the Australian Sleeve Indexed System (SIS). 
This feature helps to prevent the crossing of pipelines. Under no circumstances should these connections be altered. 
 

Alarm Test 

The user should test the alarm system for proper operation prior to each clinical use. 
 

Machine Diagnostics 

The Digital MDM Mixer performs self-checks at start up and during operation. If any errors are detected, the unit displays 
EE on the % OXYGEN display and an error code on the LPM FLOW display. If an error code appears, remove the unit from 
patient service and call Air Liquide Healthcare on 1300 36 02 02.  

To aid in the diagnostics of the unit, make note of the final error code displayed on the unit.

 

 

 

Содержание Matrx Digital MDM

Страница 1: ...y System for Porter Instrument is certified to ISO 13485 The scope of our registration is The design manufacture distribution and servicing of fluidic components and integrated systems for use in the...

Страница 2: ...r exceed the guidelines established by the National Fire Protection Association for Nonflammable Medical Gas Systems NFPA 99 WARNING During any power outage remember to turn OFF the flowmeter and manu...

Страница 3: ...third wire on the power cord when plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes or ordinances 2 The external ground lug on the rear of t...

Страница 4: ...for individual Scavenger Interfaces Monitor for N2O in the patient treatment area to ensure that controls are effective in achieving low PPM parts per million exposure TWA 25ppm in Australia Contact...

Страница 5: ...lation on page 4 Breathing Circuits Breathing circuits that can be used with any Porter and Matrx device include Porter double mask autoclavable nasal hood breathing circuit Matrx by Parker Scavenging...

Страница 6: ...he user to any problem condition within the DIGITAL MDM mixer unit or in the gas supply Displays and Indicators Figure 1 Flow Display The LPM FLOW display 4 normally displays the total gas flow rate I...

Страница 7: ...ey the total flow rate indication changes slowly for the first two steps then changes at a faster rate for the remainder of the time the key is held An audible signal indicates that a limit has been r...

Страница 8: ...ghten 2 Connect the SIS female hexagon fitting of the oxygen hose to the male SIS O2 INLET fitting on the rear of the unit see Figure 3 Hand tighten and then finish tightening with a 11 16 inch open e...

Страница 9: ...ic protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical professional Nitrous oxide although not s...

Страница 10: ...t who is awake alert and cooperative and requires relief from minimal to moderate pain and is under the continuous supervision of a healthcare professional Age Weight limitations see Relative Contrain...

Страница 11: ...ower limit while the key is held Press the flow key and hold The flow rate should increase to 9 9 LPM An audible alarm will sound at the upper limit as long as the key is held Indicated Percentage Del...

Страница 12: ...eted turn off using On Off key 8 Turn OFF the gas supply of the cylinders at the end of the day MAINTENANCE The Digital MDM Mixer is a precision analgesia mixing unit As with any medical instrument it...

Страница 13: ...centration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture up to 9 9 LPM Low end flow limited by 1 0 LPM oxygen limit Electrical 220 240 V Input Line Fr...

Отзывы: