Digi Cell Plus - Digital Oil Test Cell Information Guide
16
GB
To calibrate
or re-calibrate
OILs
> OILs <
OK
q
to select OIL TYPE to calibrate OK
CAL
Follow on-screen instructions
START
Shake cell
SAVE (or DISCARD to repeat or abort calibration)
DONE
Repeat for all OIL TYPES to be calibrated
OK
BN OIL TYPE List Name, BN Value & Calibration
DIGI+ stores calibration data for up to 8 different
OIL TYPES. In Smart DIGI+ mode stores calibration data separately for
up to 10 different
OIL TYPES (these can only be edited in DIGI_CFG.csv file on USB drive)
NOTE: It is recommended to use the oil grade or name as the “Name” during DIGI+ setup process
Action
Visual
Description
To assign a
name to OILs
> OILs <
OK
q
to select OIL TYPE OK
NAME
uq
to select character NEXT
Repeat steps above and when finished press NEXT repeatedly until
OIL TYPE NAME VALID
OK
*NOT CALIBRATED*
NOTE: OIL TYPE needs a value of BN before calibration is possible.
To edit OILs
list
> OILs <
OK
q
to select OIL TYPE to edit OK
NAME
uq
to select character NEXT
Repeat steps above and when finished press NEXT repeatedly until
OIL TYPE NAME VALID
OK
*NOT CALIBRATED*
NOTE: OIL TYPE needs a value of BN before calibration is possible.
To assign a
BN value to
OILs
> OILs <
OK
q
to select OIL TYPE to edit OK
BN
q
to scroll through numbers OK
Repeat steps above and when finished press NEXT repeatedly until
OIL TYPE NAME VALID
OK
*NOT CALIBRATED*
NOTE: OIL TYPE needs a value of BN before calibration is possible.
Power on
SETUP