Digi Cell Plus - Digital Oil Test Cell Information Guide
14
GB
Setup Menu Navigation
Setting Date & Time
>
Date/Time
<
OK
q
to select YEAR
u
q
to select MONTH
u
q
to select DAY OK
q
to select HOUR (24H)
u
q
to select MINUTES OK
Power on
SETUP
BN Calibration
Gas is generated from a chemical that reacts with available (remaining) BN additive in the sample. Gas pressure from
used oil is compared to new oil as a ratio:
Pressure is generated in DIGI+ proportional to the remaining BN additive in the used oil sample and is compared to
pressure from the new oil.
• the calibration can be checked by testing new oil in the BN, in place of used oil
• it is good practice to check and recalibrate BN, if necessary, every six months