ilization, inspect the device in the autoclave bag 
for integrity. If suspicion about the item exists, 
discard it and order a replacement from Parkell 
or your dealer.

• 

DO NOT USE DRY HEAT OR CHEMCLAVE on 

the sheath or the insert.

• Detachable handpiece sheath and ultrasonic 

inserts from Parkell are designed to withstand a 

minimum of 250 autoclaving cycles when repro-

cessed as per the recommended autoclaving 

instructions by the manufacturer.

How to Replace the Filter Disk

The water filter disk coming off of the rear hose of 

the unit should be replaced when it becomes dark 

or clogged with debris so that water flow is not 

blocked (1-3 months). Replacement Filters are 

available from Parkell (

h

D419 – 10 per package).

1. Close any water valves, and bleed all water lines 

of residual water pressure.

2. Disconnect the scaler water supply hose from 

the water supply at the quick-connect. 

3. 

 Remove the old filter disk by holding it in one 

hand and loosening it by twisting it off the hose 

connectors on either side. They rotate in oppo-

site directions.

4. 

 Install the new filter disk in the same position 

as the old one, and tighten the connectors in 

reverse.

5. Reconnect the scaler water supply hose to the 

water supply via the quick-connect and check 

for leaks.

6. 

 Note: If the filter becomes clogged and you do 

not have a replacement disk, the hose connec-

tors may be locked together TEMPORARILY 

without the filter disk to allow emergency opera

-

tion only.

Troubleshooting

 

Unit does not operate  

(“power on” indicator does not light):

• Transformer cable disconnected from wall 

outlet or scaler. Reconnect properly.

• Wall outlet not live. Reconnect properly.
• Foot Pedal or cable is damaged. 

Contact Parkell.

• Unit fuse has failed. Contact Parkell.

 

Insert does not vibrate properly: 

• Faulty, damaged or worn insert. Replace.
• Insert not correctly seated in handpiece. 

Reinsert properly.

• Use a new Parkell

®

 brand insert.

• Power control not correctly adjusted. 

Readjust.

 

Excess heat at handle or tip of insert: 

• Faulty, damaged or worn insert. Replace.
• 

Too little water flow. Insert may be blocked. 

Inspect and clear blockage, or replace insert.

• Power too high for the amount of water. 

Increase water flow.

• Excessive hand pressure applied. Correct 

technique.

 

 No water spray when foot pedal is 

depressed:

• Water line blocked or kinked. Correct.
• Water passage in the scaling insert clogged. 

Inspect and clear blockage, or replace insert.

• 

Water filter clogged. Replace.

• No water supply connected. Correct.

 

 Insert does not go into handpiece:

•  Insert stack is bent. Straighten plates 

carefully by hand and reinsert.

• O-ring is too tight. Lubricate O-ring with water 

and use twisting motion to seat insert.

• Autoclavable handpiece sheath is not 

attached to handpiece.

 

Insert falls out of handpiece or water leaks 

from front of handpiece:

• O-ring damaged or worn. Replace O-ring on 

insert and/or handpiece.

 

Water leaks at base of handpiece:

• Orange O-ring at end of handpiece is missing 

or damaged.

 

 Insert comes out of handpiece when pedal 

is depressed:

• Water pressure is too high. Adjust water 

pressure to 15-35 psi at dental unit connec-

tion or in floor junction box, or replace insert 

O-ring.

User Maintenance and 

Authorized Service

• The Parkell TurboSENSOR requires very little 

maintenance for years of trouble-free operation. 

There are no user-serviceable parts inside the 

control unit or the foot pedal. Opening the cases 

will void all warranties. Authorized Parkell person-

nel must make all internal repairs.

Содержание TurboSENSOR

Страница 1: ...the device is being connected disconnected or when not in use The power supply cord functions as the AC power disconnect device for the scaler The transformer should be located above and away from an...

Страница 2: ...5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storage Conditions 17 40 C 20 70 RH non condensing Certifications All electrical devices have been certified to IEC 606...

Страница 3: ...ontrol Knob for normal operation or by depressing the Foot Pedal to enter the Turbo Mode Use the Power Control Knob to set the initial scaling power for the procedure at hand Turn the knob in a clockw...

Страница 4: ...se of hand pressure Failure to use the lowest effective power setting for the case at hand Insufficient water flow Scaling with the point of the insert instead of the side Inserts that are damaged ben...

Страница 5: ...Reconnect properly Foot Pedal or cable is damaged Contact Parkell Unit fuse has failed Contact Parkell Insert does not vibrate properly Faulty damaged or worn insert Replace Insert not correctly seate...

Страница 6: ...era ture of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must...

Страница 7: ...erbro chen werden wenn das Ger t angeschlossen oder getrennt wird oder wenn es nicht in Gebrauch ist Das Stromversorgungskabel fungiert als Wechsel strom Trenneinrichtung f r den Scaler Der Transforma...

Страница 8: ...ierbare H lsen eine am Handst ck befestigt hD576 2 Wasserleitungsfilter eineranWasserleitungangebracht 1 Transformator mit Kabel 1 Bedienerhandbuch Gebrauchsanleitung EINS TZE Magnetostriktive Eins tz...

Страница 9: ...r Leistungsreglerknopf auf die Einstellung Perio gestellt ist leuchtet die Perio LED auf die Spitze vibriert dann bei Bet tigen des Fu pedals mit wesentlich niedrig erer Amplitude als ein herk mmliche...

Страница 10: ...IEREN da es durch die Hitze beim Autoklavieren zerst rt wird Ablagerungen auf dem Steuerger t soll ten mit einem mit Wasser befeuchteten Papiertuch entfernt werden F r die Desinfektion kann ein Papier...

Страница 11: ...itigen ggf die Blockade oder tauschen Sie den Einsatz aus Zu hohe Leistung f r die Menge Wasser Erh hen Sie den Wasserfluss berm iger Druck mit der Hand angewendet Achten Sie auf korrekte Durchf hrung...

Страница 12: ...raturbetrieb durchgef hrt werden Wenn Sie Fragen zu oder Probleme bei der Installation oder Anwendung Ihres TurboSENSOR Scalers haben sollten rufen Sie unseren Technical Support Service unter 1 800 24...

Отзывы: