background image

4

• Protect patient’s eyes, lips, cheek, tongue or other 

vulnerable soft tissues when using this device. 

Inadvertent contact with Insert tip may cause a 

slight burn.

•   Inserts and handpiece sheaths must be sterilized 

before each use.

• Use of a face mask by the clinician, and high-vol-

ume intraoral suction, is strongly recommended 

when operating this device to avoid inhalation of 

contaminated aerosols.

• In order to prevent a possible slight burn to the 

patient or clinician, 

NEVER OPERATE UNIT 

WITHOUT WATER FLOWING.

• Use the lowest effective scaling power for  

the case at hand. This keeps heat generation to 

a minimum.

• Before using the insert in the oral cavity, adjust 

the water spray following the graphics on the face-

plate of the scaler over a sink or cuspidor until 

desired mist is obtained. For maximum patient 

comfort, use a good flow of water as a coolant, 

lubricant and to flush out debris.

• If patient is new to ultrasonic scaling, explain to 

them what to expect. Usually, the patient should 

not experience discomfort. If patient experiences 

uncomfortable heat, adjust power and water 

controls accordingly.

• 

 Do not test a scaler tip for vibration on your fingers 

while the unit is operating. This is not a valid test 

of how scaling feels to the patient.

• Hold the handpiece in a comfortable pen-grasp. 

To remove deposits from teeth, always use a 

light brushing stroke with the side of the insert 

tip, in contact with and parallel to the root 

surface. Excess pressure will not improve scal-

ing action, but may cause heating and pain to the 

patient. Keep the scaler insert in motion at all 

times, using repeated gentle strokes to remove 

all tenacious deposits.

• In the event that the patient experiences discom-

fort due to the sound and vibration from the Scal-

ing procedure, provide ear protection.

• Do not use the point of the insert on the tooth 

surface. You may gouge the tooth.

• Pause occasionally during scaling by removing 

foot from pedal to evaluate deposit removal.

•  Exercise caution near ceramic restorations, as 

they can discolor or fracture if stressed.

Common Operator Errors That 

May Result in Poor Clinical 

Performance

• Failure to keep insert tip parallel to the long axis 

of the tooth.

• Excessive use of hand pressure.

•    Failure to use the lowest effective power setting 

for the case at hand.

• 

  Insufficient water flow.

• Scaling with the point of the insert instead  

of the side.

•   Inserts that are damaged, bent or worn out.

Cleaning and Infection Control

• Consult www.CDC.gov for the “Guidelines for 

Infection Control in Dental Health-Care 

Settings” and “Statement on Reprocessing 

Dental Handpieces”.

• DO NOT AUTOCLAVE the TurboSENSOR 

control unit, as autoclave heat will destroy it. The 

control unit should be cleaned of debris with a 

water-damp paper towel, disinfected by wiping 

with a paper towel that has been saturated with 

an EPA-registered low-level (HIV/HBV claim) to 

intermediate-level (tuberculocidal claim) hospi-

tal disinfectant, wiped of chemical residue with 

a water-damp paper towel, and dried. Utilize the 

surface disinfecting protocol of the disinfectant 

manufacturer. 

• Autoclaving does not remove debris that has 

accumulated on the sheath or the insert. Failure 

to adequately remove debris will result in inade-

quate sterilization. The sheath and ultrasonic 

insert must be washed in soap and water either 

manually or in an ultrasonic cleaner, and rinsed in 

running water prior to bagging them for process-

ing in a steam autoclave. Rinse the Sheath under 

warm running water for 30 seconds to remove any 

external or internal soil or debris. Using a soft 

soapy cleaning brush to assist in the cleaning, if 

necessary. Use non-ammoniated detergent or 

dishwashing soap. Do not use ammoniated 

cleansers or disinfectants. Rinse the Sheath 

again under warm running water for 30 seconds 

to remove any residual soap and blot dry with a 

dry lint-free towel. Wipe Insert with a lint-free 

towel that has been saturated in an EPA-ap-

proved, hospital-grade intermediate or high-level 

disinfectant (minimum 2.5% Glutaraldehyde solu-

tion), following the instructions for use provided 

by the disinfectant manufacturer. Dry Sheath with 

a dry lint-free towel.

• Sheaths and inserts may be sterilized in any 

conventional steam autoclave following manu-
facturer’s instructions. A typical steam steriliza-
tion cycle is 132 +/ 2°C for 4 mins. (Vacuum) or 
132 +/ 2°C for 15 mins. (Gravity), followed by a 
15 minute minimum cool-down period. Use a 
steam sterilization pouch that is compliant with 
ISO 11140-1 Type 4 and ISO 11607-1 standards. 
Once items are sterilized, adhere to the shelf-life 

specified by the pouch manufacturer. After ster

-

Содержание TurboSENSOR

Страница 1: ...the device is being connected disconnected or when not in use The power supply cord functions as the AC power disconnect device for the scaler The transformer should be located above and away from an...

Страница 2: ...5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storage Conditions 17 40 C 20 70 RH non condensing Certifications All electrical devices have been certified to IEC 606...

Страница 3: ...ontrol Knob for normal operation or by depressing the Foot Pedal to enter the Turbo Mode Use the Power Control Knob to set the initial scaling power for the procedure at hand Turn the knob in a clockw...

Страница 4: ...se of hand pressure Failure to use the lowest effective power setting for the case at hand Insufficient water flow Scaling with the point of the insert instead of the side Inserts that are damaged ben...

Страница 5: ...Reconnect properly Foot Pedal or cable is damaged Contact Parkell Unit fuse has failed Contact Parkell Insert does not vibrate properly Faulty damaged or worn insert Replace Insert not correctly seate...

Страница 6: ...era ture of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must...

Страница 7: ...erbro chen werden wenn das Ger t angeschlossen oder getrennt wird oder wenn es nicht in Gebrauch ist Das Stromversorgungskabel fungiert als Wechsel strom Trenneinrichtung f r den Scaler Der Transforma...

Страница 8: ...ierbare H lsen eine am Handst ck befestigt hD576 2 Wasserleitungsfilter eineranWasserleitungangebracht 1 Transformator mit Kabel 1 Bedienerhandbuch Gebrauchsanleitung EINS TZE Magnetostriktive Eins tz...

Страница 9: ...r Leistungsreglerknopf auf die Einstellung Perio gestellt ist leuchtet die Perio LED auf die Spitze vibriert dann bei Bet tigen des Fu pedals mit wesentlich niedrig erer Amplitude als ein herk mmliche...

Страница 10: ...IEREN da es durch die Hitze beim Autoklavieren zerst rt wird Ablagerungen auf dem Steuerger t soll ten mit einem mit Wasser befeuchteten Papiertuch entfernt werden F r die Desinfektion kann ein Papier...

Страница 11: ...itigen ggf die Blockade oder tauschen Sie den Einsatz aus Zu hohe Leistung f r die Menge Wasser Erh hen Sie den Wasserfluss berm iger Druck mit der Hand angewendet Achten Sie auf korrekte Durchf hrung...

Страница 12: ...raturbetrieb durchgef hrt werden Wenn Sie Fragen zu oder Probleme bei der Installation oder Anwendung Ihres TurboSENSOR Scalers haben sollten rufen Sie unseren Technical Support Service unter 1 800 24...

Отзывы: