4. Laying power cable

 Raise  the  bed  to  the highest position.
 Route  the  power  cable  on the side where

 

the user does

not get on or off the air mattress and

 

the

 

bed side table

is not used.

WARNING 

z

 

Route the power cable on the side where the user does not

 

get on or off

the air mattress and the over bed table is not used. There is a

 

risk of

tripping over the power cable and falling over. There is also a risk

 

that the

power cable may be damaged, causing electric shock or fire.

z

 

If the bed has a height adjustment function, route the power cable while

the bed has been raised to the highest position. If the power cable is not
routed at the highest position of the bed, the power cable may be pulled
and disconnected when the bed is raised, possibly causing a fire.

z

Do not route the power cable inside or under the bed. If the power cable

get caught 

in

 the casters or the mov

able

 parts of the bed, the power cable

may be damaged, 

possible 

causing electric shock or fire.

5. Connecting to the power supply

WARNING

z

 

Do not plug the power cord into a multiple-socket

outlet. Plugging the power cord into a

multiple-socket outlet or power strip shared by

other electrical equipment could cause the outlet

to exceed its rated capacity, causing the power

cable or plug to overheat, resulting in fire.

z

 

Do not plug the power cord into an outlet shared

by other ME devices. There is a risk of fire

caused by heating up of the power cable or

power plug.

z

 

Make sure to plug the power cord into an outlet

equipped with a circuit protector.

z

 

There is a risk of electric shock.

z

 

Each air mattress should be plugged into a separate,

appropriately wired, single-phase power

 

outlet of 240

VAC,10 A or greater and the power plug

 

must be

properly inserted into the outlet.

z

 

Disconnecting the power plug from the outlet can

separate the air mattress from the power.

z

 

Do not place the air mattress or other

devices in locations that hinder access
to the power plug. Doing so prevents the
power plug being disconnected
immediately when necessary.

No
te!
 

Power plug

P

ower outlet

Power cable

Содержание cococia Fuka

Страница 1: ...information about any serious incident related to this air mattress that occurs within the EU to the manufacturer and the regulating authority of the member country of residence Serious incident mean...

Страница 2: ...PRODUCTS 30 PRODUCT SYMBOL DEFINITIONS 31 AFTER SALES SERVICE 32 INTENDED USE J cococia F ka is an air mattress with an external pump which is designed to disperse body pressure when used for care in...

Страница 3: ...at the product logo faces up and the blowing tube is at the foot end of the user Head end Foot end CPR valve CPR label Top Plug Product logo Warning label Snap button Product identification label Blow...

Страница 4: ...on label Operation Panel Enlarged View Power button Status display Back raised mode button Max inflate mode button Hardness adjustment button Alternate pressure button Handle Blowing tube insertion po...

Страница 5: ...ction manual store it in a place that can be easily accessed for later reference About the warning CPR and product ID labels zWarning and precautions requiring special attention are described on label...

Страница 6: ...sk of unexpected gaps being created and part of the body getting caught resulting in injury J When getting in and out of the bed air mattress or taking a sitting position be careful not to fall or top...

Страница 7: ...Take care when handling the pump zThe air mattress may malfunction causing injury failure or damage In addition there is a risk of electric shock or fire due to a short circuit x Do not step on press...

Страница 8: ...stallation or improper connection or fixing may cause malfunction or result in an inability to supply air so the body pressure may not be dispersed effectively possibly causing bedsores zImproper fixi...

Страница 9: ...cted service life of t o ears o maintain the condition of the pump have the pump serviced regularl according to the schedule recommended o not use unapproved accessories or attempt to modif disassembl...

Страница 10: ...replace it with a new one For the replacement procedure contact your distributor see Contact Information on p 32 zIf the inside of the air mattress is stained or damaged contact your distributor see C...

Страница 11: ...OFF When ON is set the mode is set to OFF automatically after 60 minutes elapse WARNING zAdjust the usage time according to the user s medical condition If the internal pressure of the air cells cont...

Страница 12: ...p and the blowing tube is at the foot end of the user zCheck that the CPR valve plug points to CLOSE If it points to OPEN turn the plug to the CLOSE position WARNING zPlace the air mattress directly o...

Страница 13: ...sides of the coupling inward from both sides and insert into the blowing tube insertion port of the pump zInsert until you hear a click Check that both hooks on the left and right and the insertion li...

Страница 14: ...the pump Do not spill water on the pump or touch it with wet hands zWhen installing the pump do not bend the blowing tube Sufficient air may not be supplied possibly causing bedsores zTake care when...

Страница 15: ...ible causing electric shock or fire 5 Connecting to the power supply WARNING z Do not plug the power cord into a multiple socket outlet Plugging the power cord into a multiple socket outlet or power s...

Страница 16: ...ble of the pump to the outlet and hold down the power button for two seconds The pump starts working and the air cells expand Adjust the hardness of the air mattress according to the weight of the use...

Страница 17: ...ing of each function J Hold the CPR valve at the head end of the air mattress see the figure in 2 How to lay the sheet and turn the plug from CLOSE to the OPEN position zWhen supplying air to the air...

Страница 18: ...nants blood feces etc remove them Use soapy water commercially available soap such as neutral detergent diluted with water to wipe the surface of the cover on the air mattress to remove stains CAUTION...

Страница 19: ...the cover and therefore do not need to be maintained frequently zWhen maintaining the air cells do not separate each air cell from the air mattress Remove the coupling from the pump and discharge air...

Страница 20: ...handling method of each disinfectant z76 9 81 4 z70 z0 05 0 2 z0 05 0 2 z0 05 Ethanol for disinfection Isopropanol benzalkonium chloride benzethonium chloride chlorhexidine gluconate Replacing the fil...

Страница 21: ...ection J For safe use inspect the parts at least once a year The parts will exhibit wear and deterioration depending on the frequency of use and the environment Regularly inspect the following items J...

Страница 22: ...e pump Is the CPR valve set to OPEN Turn the plug of the CPR valve to the CLOSE position Air is not present in some air cells Is the alternate pressure lamp ON Since the alternate pressure is set to O...

Страница 23: ...the pump Is the pump hooked onto a part that rattles Hook the pump onto a part that does not rattle or place it Contacting the bottom Is the back section base of the bed raised while the back raised...

Страница 24: ...ward from both sides and remove the coupling from the pump CAUTION zWhen removing the coupling from the pump press the levers on both sides of the coupling and pull the coupling out of the pump Do not...

Страница 25: ...educe the odor zKeep the instruction manual in a safe place so that it does not become lost CAUTION zStore the product within the specified storage temperature humidity and air pressure ranges Failure...

Страница 26: ...Size cm Width 25 Depth 11 Height 22 Material Case ABS resin flame resistant chemical resistant Power supply Plug 2P Length 4 6 m Product weight kg 2 2 Guaranteed conditions Operating temperature humid...

Страница 27: ...her than those specified and sold by the manufacturers of Cococia Fuka are used as replacement parts for internal components it may result in an increase in emission or a decline in immunity of Cococi...

Страница 28: ...trengths from fixed RF transmitters as determined by an electromagnetic site survey must be less than the compliance level in each frequency range Do not use the pump for Cococia F ka and portable and...

Страница 29: ...itter in watts W according to the transmitter manufacturer NOTE1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE2 These guidelines may not apply in all situat...

Страница 30: ...ommunication devices are used The electromagnetic interference between the RF communication device and the pump for Cococia F ka can be prevented by maintaining the minimum distance between the two as...

Страница 31: ...ombination check the instruction manuals for beds bed side rails and swing arm bed grips side grips and Paramount Bed s catalogs and select suitable products that meet the following suitability standa...

Страница 32: ...lease request for repair work Turns o automatically in 120 minutes Lit up ON Max in ate mode selection Turns o automatically in 60 minutes Lit up ON Be careful not to get a body part caught in the gap...

Страница 33: ...unt Bed keeps replacement parts necessary to maintain stretcher performance in stock for at least 6 years after the discontinuation of production 3 Questions regarding after sales service Contact us s...

Страница 34: ...MEMO 33...

Страница 35: ...MEMO 34...

Страница 36: ...Original Instruction Manual Made by TorindoCo Ltd 2020 11 AL300429 V1 01...

Отзывы: