DigitalFlow™ GC868 Startup Guide (1 and 2-Channel)
1-16
Chapter 1. Installation
The spacing should now appear similar to that in Figure 1-20 below.
Figure 1-20: CFG-PI Fixture with Calculated Spacing
1.4.5.4
Installing the
Transducers
1.
Check to be sure the second CFG-PI holder bracket is correctly positioned.
2.
Apply a bead of CPL-16 couplant 6 mm (0.25 in.) wide on each transducer face
(see Figure 1-21 below).
Figure 1-21: Couplant on Transducer Face
Note:
Do not slide the transducer with couplant along the surface of the pipe when
mounting the transducer.
3.
With one hand, mount one transducer into the PI fixture. With the other hand,
tighten the thumbscrew, gradually pushing the transducer down to the pipe
surface. Use a wrench to tighten the backing nut to prevent loosening due to
vibration and thermal expansion.
IMPORTANT:
Do not use pliers or a wrench on the thumbscrew.
4.
Repeat step 3 for the other transducer.
S
S
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 44: ...Chapter 1 Installation 1 33 DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 48: ...Chapter 1 Installation DigitalFlow GC868 Startup Guide 1 and 2 Channel 1 45 no content intended for this page...
Страница 58: ...Chapter 2 Initial Setup 2 9 DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 60: ...Chapter 2 Initial Setup DigitalFlow GC868 Startup Guide 1 and 2 Channel 2 11 no content intended for this page...
Страница 72: ...Chapter 4 Specifications 4 5 DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...