DigitalFlow™ GC868 Startup Guide (1 and 2-Channel)
1-28
Chapter 1. Installation
1.8.4.3
Wiring the Ethernet
Interface
A modified GC868 can use the Ethernet interface to communicate to an internal
network. An optional Ethernet card with a unique MAC (IP) address (installed only in
slots 5 or 6) includes an RJ45 connector. To connect the Ethernet-enabled GC868 to
the network, insert the jack of an RJ45 cable into the RJ45 connector, route the cable
through the bottom of the GC868, and wire the other end of the cable to the Ethernet
network according to the manufacturer’s instructions. An external connection is
required between the Ethernet option card and the GC868’s RS232 connector, as
shown in Table 1-5 below.
Note:
The MAC address for a specific GC868 is included with customer
documentation. For more information on setting up the MAC address, refer
to Chapter 6 of the
Programming Manual.
1.8.4.4
Wiring the
MODBUS/TCP
Interface
Customers can also use a modified GC868 that provides a MODBUS/TCP interface to
communicate to an internal network. An optional MODBUS/TCP card with a unique
MAC (IP) address (installed only in slots 5 or 6) includes an RJ45 connector. To
connect the MODBUS/TCP-enabled GC868 to the network, insert the jack of an RJ45
cable into the RJ45 connector, route the cable through the bottom of the GC868, and
wire the other end of the cable to the Ethernet network according to the
manufacturer’s instructions.
Note:
The MAC address for a specific GC868 is included with customer
documentation. For more information on setting up the MAC address, refer
to Chapter 6 of the Programming Manual.
1.8.4.5
Wiring the
Foundation
Fieldbus Network
Fieldbus network connections are made at J8/J9, pins 1 and 2 (see Figure 1-37
below). Optionally, a shield can be connected to J8/J9
pin 3, depending on the network wiring. Connector J8 or J9 will be installed
depending on the option ordered by the customer.
No connections are made to J8/J9, pins 7 and 9, under normal operation. If it is
desired to reset the network board to factory defaults:
1.
Connect a jumper between J8/J9 pin 7 and pin 9.
2.
Power cycle the instrument.
3.
Ten seconds after the power has been restored to the unit, remove the jumper to
return the network board to normal operation.
Table 1-5: RS232 to Ethernet Interconnections
GC868 Type
Terminal Block
Terminal Block
Wall Mount
RS232 on Main Board TB1 on Ethernet Card
TX
Pin 1
RX
Pin 2
RTN
Pin 3
Rack Mount
RS232 on Main Board TB2 on Ethernet Card
TX
Pin 1
RX
Pin 2
RTN
Pin 3
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 44: ...Chapter 1 Installation 1 33 DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 48: ...Chapter 1 Installation DigitalFlow GC868 Startup Guide 1 and 2 Channel 1 45 no content intended for this page...
Страница 58: ...Chapter 2 Initial Setup 2 9 DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...
Страница 60: ...Chapter 2 Initial Setup DigitalFlow GC868 Startup Guide 1 and 2 Channel 2 11 no content intended for this page...
Страница 72: ...Chapter 4 Specifications 4 5 DigitalFlow GC868 Startup Guide 1 and 2 Channel no content intended for this page...