Cyanuric Acid 

Range: 2 – 200 mg/l  

Wavelength – 530 and 575 nm 

Method – Turbidity  

1   Fill test tube with sample to the 10 ml mark. 

2

Add one Cyanuric Acid tablet and allow to disintegrate 

for at least 

two minutes

.  

A cloudy solution indicates the presence of cyanuric 

acid. 

4   Crush  any  remaining  undissolved  tablet  and  mix  to 

ensure uniformity. 

Take the photometer reading.  

 

Reagents 

AP 087 – Cyanuric Acid Tablets 

 

 

 

 

 

 

 

19

 

Alkalinity (Total) 

Range: 10 – 500 mg/l  

Wavelength – 575 nm 

Method – Acid and Indicator 

Fill test tube with sample to the 10 ml mark. 

2  

Add one Alkaphot tablet, crush and mix until all the 

particles  have  dissolved.    It  is  important  to  ensure 

that  all  of  the  tablet  particles  have  dissolved 

otherwise the result may be inaccurate. 

Stand for 

one minute

4  

Take 

the 

photometer 

reading. 

The 

reading 

represents  the  total  alkalinity  of  the  sample  as 

milligrams per litre (parts per million) CaCO

3

 

Reagents 

AP 188 – Alkaphot Tablets 

 

Notes 

For accurate results with Alkaphot, crush the tablet very 

thoroughly,  stand  for  one  minute  and  then  remix  the 

solution.   Look  carefully at the bottom of the tube.  If a 

thin  yellow  layer  forms,  then  mix  the  tube  again.  This 

ensures  that  reaction  is  complete.  The  result  should not 

change on standing. 

 

Содержание Pooltest 3

Страница 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Страница 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Страница 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Страница 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Страница 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Страница 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Страница 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Страница 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Страница 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Страница 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Страница 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Страница 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Страница 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Страница 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Страница 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Страница 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Страница 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Отзывы: