palintest Pooltest 3 Скачать руководство пользователя страница 14

 

 

28

The  colour  range  of  the  phenol  red  test  is  yellow, 

through  orange,  to  red.  The  formation  of  an  intense 

purple  coloration  shows  that  the  indicator  has  been 

affected by high chlorine or bromine residuals. In such 

cases the result should be disregarded. 

Cyanuric Acid

:  

The  range  of  the  cyanuric  acid  (CNA)  test  is  2  -  200 

mg/l.  Higher  levels  can  be  tested  by  first  diluting  the 

sample 

with 

mains 

or 

deionised 

water 

and 

then 

applying  the  appropriate  dilution  factor.  Palintest’s 

Dilution 

Tube  (PT 512), 

which  is  available  as  an 

optional extra, should be used for the dilution. 

Calcium Hardness

:  

The  expression  of  hardness  results  sometimes  causes 

confusion.  It  is  normal  practice  to  express  the  results 

of  hardness  tests  as  mg/l  CaCO

3

  (calcium  carbonate). 

This  is  merely  a  convention  to  allow  the  comparison  of 

different  results  and  does  not  necessarily  indicate  that 

the hardness is present in the water in this form. 

Results  may  also  be  expressed  in  literature  as  mg/l  Ca.  

The  instrument  will  not  do  this  calculation  automatically 

but to convert mg/l CaCO

3

 to mg/l Ca multiply by 0.4. 

 

 

29

Bromine

:  

For  most  purposes  it  is  sufficient  to  simply  measure  the 

total  bromine  residual  (using  DPD  No  1  tablets)  since 

both free and combined bromine are active disinfectants. 

Note  that  a  too  high  bromine  level  (above  20  mg/l)  can 

cause  bleaching  of  the  pink  colouration  formed  in  the 

DPD  test  and  give  a  false  negative  or  low  result.  If  a 

colourless  or  weakly  coloured  test  solution  is  obtained 

when  bromine  is  known  to  be  present,  check  for  the 

possibility of bleaching by repeating the test on a sample 

diluted with bromine/chlorine-free water. 

Water Balance: 

In  swimming  pools,  the  effects  of  the  pH,  hardness  and 

alkalinity  are  all  inter-related.    The  likelihood  of  corrosive 

or  scale  problems  can  be  predicted  using  the  Palintest 

Water  Balance  Calculation.    The  resulting  value  indicates 

to  users  the  corrosion  or  scale  forming  tendency  of  the 

water. The Palintest method is a simplified version of the 

Langelier  Index  and  this  can  be  calculated  from  the 

Palintest result by subtracting 11.1. 

Calculation

1

Take  a  sample  of  the  water  and  measure  the  calcium   

hardness.    Using  the  table  overleaf,  read  off  the  factor 

which corresponds nearest to the calcium hardness value 

measured. Record this Calcium Hardness Factor (Line 1). 

Содержание Pooltest 3

Страница 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Страница 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Страница 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Страница 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Страница 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Страница 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Страница 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Страница 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Страница 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Страница 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Страница 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Страница 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Страница 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Страница 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Страница 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Страница 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Страница 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Отзывы: