INSTRUMENT  

For an illustration on the instrument layout, see Section 1. 

For information on performing a test, see Section 2. 

 

Replacing the Batteries 

Replace the battery when the 

 symbol remains on the 

display.  Use  2  x  1.5v  alkaline  ‘AA’  batteries,  MN  1500, 

LR6,  E91,  AM3  or  equivalent.    Remove  batteries  from 

instrument  if  it  is  to  be  stored  or  left  unused  for  a  long 

period of time. 

 

Liability 

Under  no  circumstances  shall  Palintest  Ltd  be  liable  for 

loss  of  life,  property,  profits  or  other  damages  incurred 

through the use or misuse of their products. 

Disposal 

Waste Electrical and Electronic Equipment (WEEE). 

Natural  resources  were  used  in  the  production  of  this 

equipment.  This  equipment  may  contain  materials  that 

are hazardous to health and the environment.  

 

 

 

23

To  avoid  harm  to  the  environment  and  natural  resources, 

the use  of appropriate recycling systems  is recommended. 

The  crossed  out  wheeled  bin  symbol  on  the  meter 

encourages  you  to  use  these  systems  when  disposing  of 

this equipment. 

 

Error Messages 

The photometer will display an error message in the unlikely 

event  of  malfunction.    These  error  messages  are  mainly 

designed to assist service staff in diagnosing instrument faults.  

In  the  event  of  an  error  message  appearing  on  the 

photometer  display,  contact  your  local  Palintest  Technical 

Services Department or your local distributor. 

Error  messages  are  coded  7,  8  and  9  and  all  relate  to 

blanking  the  instrument.    In  the  first  instance,  the  user 

should 

check 

the 

operating 

technique 

and 

sample 

clarity.    If  these  are  in  order,  then  these  errors  indicate 

a fault in the optics :- 

Error 7 indicates too much light  – remove the instrument 

from bright light. 

Error 

indicates 

fault 

with 

one 

of 

the 

optics 

components, and requires service assistance. 

Error  9  indicates  not  enough  light  –  follow  ‘Cleaning  the 

Optics’ routine. 

If  the  problem  persists,  contact  your  local  Palintest 

branch or distributor. 

Содержание Pooltest 3

Страница 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Страница 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Страница 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Страница 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Страница 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Страница 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Страница 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Страница 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Страница 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Страница 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Страница 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Страница 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Страница 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Страница 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Страница 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Страница 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Страница 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Отзывы: