14

Chlorine 

Range 

Cl

2

 5    : 0.01 –  5 mg/l  

Cl

10 : 0.01 – 10 mg/l  

Wavelength – 530 and 575 nm 

 

Method – DPD 

Free Chlorine 

1

Rinse  test  tube  with  sample  leaving  a  few  drops  in 

the tube. 

2

Crush  the  DPD  No  1  tablet  (or  DPD-XF  for  high 

range  instruments)  in  two  or  three  drops  of  the 

water sample until the tablet is thoroughly crushed.  

3

Add  the  10  ml  test  solution,  mix  and  seal  the  tube 

with the cap. 

4

Gently  invert  the  tube  to  remove  any  bubbles  from 

the inner walls of the tube. 

5

Take photometer reading.  

6

Retain  test  solution  if  total  chlorine  test  required 

(see overleaf for further instructions). 

 

 

 

 

 

 

15

Total Chlorine 

Carry  out  this  test  on  the  solution  remaining  from 

the Free Chlorine test. 

If  any  non-chlorine  shock  treatment  chemicals  have 

been added to the pool, see additional notes below. 

Add  one  DPD  No  3  tablet  (or  DPD-XT  for  high 

range instruments), crush and mix to dissolve. 

Stand for 

two minutes

Take  photometer  reading.    Ensure  the  display 

shows the test range that you desire. 

 

To  obtain  COMBINED  CHLORINE  residual,  subtract 

Free Chlorine result from Total Chlorine result :- 

ie Combined Chlorine = Total Chlorine - Free Chlorine 

 

For notes on chlorine testing, see overleaf. 

 

Notes 

If any non-chlorine shock treatment chemicals have 

been  added  to  the  pool,  add  one  DPD  Oxystop  (AP 

017)  tablet,  crush  and  mix  to  dissolve  prior  to 

adding  the  DPD  No  3  tablet  to  the  test  solution.  

Stand  for  one  minute  before  proceeding.    This  will 

prevent  a  response  caused  by  the  shock  treatment 

chemicals. 

Содержание Pooltest 3

Страница 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Страница 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Страница 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Страница 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Страница 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Страница 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Страница 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Страница 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Страница 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Страница 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Страница 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Страница 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Страница 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Страница 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Страница 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Страница 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Страница 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Отзывы: