Calcium Hardness 

Range: 5 – 500 mg/l  

Wavelength – 575 nm 

Method – Indicator 

1

Fill test tube with sample to the 10 ml mark. 

2

Add one Calcicol No 1 tablet, crush and mix to dissolve. 

3

Add one Calcicol No 2 tablet, crush and mix to dissolve. 

4

Stand for 

two minutes

5

Take the photometer reading. 

Reagents 

AP 252 – Calcicol No 1 and No 2 Tablets 

Notes 

The  Calc  NaCl  setting  is  for  salt  pools  only.    This 

requires 

different 

calibration 

but 

the 

method 

outlined for the standard Calcicol test is still valid. 

Magnesium hardness (up to 200 mg/l as CaCO

3

) does 

not interfere with the test. 

Iron at levels above 10 mg/l may cause low results. Zinc 

above 5 mg/l may cause high results. 

 

 

21

Bromine 

Range: 0.02 – 10 mg/l  

Wavelength – 530 nm 

Method – DPD 

Rinse  test  tube  with  sample  leaving  a  few  drops  of 

sample in the tube. 

Crush  the  DPD  No  1  tablet  in  two  or  three  drops  of 

the  water  sample  until  the  tablet  is  thoroughly 

crushed. 

Add  the  10  ml  test  solution,  mix  and  seal  the  tube 

with the cap. 

Gently  invert  the  tube  to  remove  any  bubbles  from 

the inner walls of the tube 

Take the photometer reading.

 

 

Reagents 

AP 011 – DPD No 1 Tablets 

Содержание Pooltest 3

Страница 1: ...lts for the testing of water Our instruments are of the highest quality and fully waterproof Palintest s experience built up over the last 50 years is the reason why our instruments and reagents are u...

Страница 2: ...Rods Test Tube Brush Photometer Tubes Reagents Chlorine Free and Total pH Phenol Red Cyanuric Acid Alkalinity Calcicol Bromine For re ordering codes and accessories please see Section 6 If you have b...

Страница 3: ...easurement between test samples and colour standards a colorimeter can be used for quantitative measurement of the amount of coloured light absorbed by a sample with reagents added in reference to a u...

Страница 4: ...l quantities of red light pass through the solution 9 Blanks and Samples Palintest photometers use a BLANK tube to set the instrument to blank and a SAMPLE tube to take the reading A BLANK tube is a t...

Страница 5: ...drips or condensation before placing in the photometer 5 Do not leave tubes standing in the photometer test chamber Remove the tubes immediately after each test 6 Immediately wipe up any drips or spil...

Страница 6: ...1 5 mg l range Cl2 10 Chlorine Free and Total 0 01 10 mg l range pH pH pH Units CNA Cyanuric Acid Alk T Alkalinity Total Calc Calcium Hardness Calc NaCl Calcium Hardness Salt Water Pools Only Br Bromi...

Страница 7: ...ion remaining from the Free Chlorine test 2 If any non chlorine shock treatment chemicals have been added to the pool see additional notes below 3 Add one DPD No 3 tablet or DPD XT for high range inst...

Страница 8: ...cal Palintest branch distributor or see www palintest com Reagents AP 011 DPD No 1 Tablets Low Range AP 013 DPD XF Tablets High Range AP 031 1 DPD No 3 Tablets Low Range AP 033 1 DPD XT Tablets High R...

Страница 9: ...sample to the 10 ml mark 2 Add one Alkaphot tablet crush and mix until all the particles have dissolved It is important to ensure that all of the tablet particles have dissolved otherwise the result...

Страница 10: ...standard Calcicol test is still valid Magnesium hardness up to 200 mg l as CaCO3 does not interfere with the test Iron at levels above 10 mg l may cause low results Zinc above 5 mg l may cause high r...

Страница 11: ...ng systems is recommended The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment Error Messages The photometer will display an error messa...

Страница 12: ...ight transmission and affect readings To clean the optics gently clean the internal surfaces of the optics with a soft non abrasive cloth Do not use solvents Deposits may be removed with a slightly da...

Страница 13: ...S are used to oxidise organic contaminants in the water and regenerate free chlorine from chloramines Typical products are Oxybrite Oxysure and Purolyte Plus 27 Excess MPS reacts in the DPD test It do...

Страница 14: ...mg l Ca multiply by 0 4 29 Bromine For most purposes it is sufficient to simply measure the total bromine residual using DPD No 1 tablets since both free and combined bromine are active disinfectants...

Страница 15: ...ness Alkalinity as mg l CaCO3 Factor 20 1 0 200 2 0 30 1 2 250 2 1 40 1 3 300 2 2 50 1 4 350 2 25 66 1 5 400 2 3 80 1 6 500 2 4 90 1 65 550 2 45 100 1 7 600 2 5 125 1 8 700 2 55 150 1 9 800 2 6 31 5 C...

Страница 16: ...the European CE Mark of Compliance for electro magnetic compatibility EMC 33 Sample Dilution and Collection When the test result is outside the concentration range of the test the photometer will disp...

Страница 17: ...087 AP 087 Alkaphot PM 188 AP 188 Calcicol PM 252 AP 252 EDTA Tablets AT 090 DPD Oxystop AP 017 Accessories Description Product Code Test Tube Brush PT 663 Replacement Photometer Tubes x5 PT 555 Photo...

Отзывы: