Remove the intake screen by pulling down with sufficient force to unseat the fastening
clips from the mating surface. Scrub debris from the intake screen using a stiff brush.
Rinse with water. Verify all holes are unplugged before replacing the screen on the pump
intake.
9.6: Refilling the Tank
Refill the tank and add cleaning solution as described in sections 9.1 and 9.2.
9.7: Cleaning
The SCA should only be cleaned with mild soap and water with a sponge or cloth. Never
immerse the system or use a spray nozzle or hose to clean it. Any liquid in the electronics
area of the system may damage the system and void the warranty. Before using any
cleaning or decontamination method except those recommended by the manufacturer,
users should check with the manufacturer that the proposed method will not damage the
equipment.
9.8: Moving
Always use two people to lift the SCA. The SCA should always be drained before moving.
To lift, have each person stand on either side of the unit and place their hands into the
indentations under the unit. Lift the unit to a cart or wheeled table for transport.
Figure 14: Lifting the Unit
Figure 13 Removing pump intake screen