1: Site Preparation
1.1: Locating the Unit
•
Place the SCA on a sturdy horizontal surface or cart capable of supporting 200 lbs
(the weight when filled with water).
•
Ensure there is no Aluminum or Zinc in the area where the SCA will be used. THE
CLEANING SOLUTION CONTAINS SODIUM HYDROXIDE AND REACTS
VIOLENTLY WITH ALUMINUM OR ZINC.
•
The selected location should be isolated from unintentional contact from those
working in the area.
•
It should be near a grounded wall outlet such that the power cord does not pose a
hazard to people or equipment passing by.
•
The power cord plug-in receptacle is a disconnecting device that should be easily
accessible at all times. Position the unit so that it can be easily reached.
•
It is strongly recommended that the unit be placed on a circuit with its own breaker
with a GFCI.
•
Place the unit in a well-ventilated area.
1.2: Preparations for Draining and Disposal
•
Placing the unit near a drain or sink will make draining the unit much easier.
Alternatively, the SCA can be kept on a cart for easy transport to a sink or drain.
•
Use a 1” inner diameter hose made of PVC or equivalent on the drain tube to
facilitate draining. Ensure the tubing or hose can withstand the temperature and
caustic properties of the cleaning solution.
Used solution must be disposed of under applicable local waste disposal regulations.
It is
the responsibility of the user to determine and verify what the local disposal
regulations are and to follow those regulations.