SECTION 4
OPERATION
50
4.5.2.1 TIME BASED TRANSIENT EXAMPLE
The following is an example of how to create a MIL-STD-704D time based transient. (See section 4.4.2
regarding pre-stored transients)
Start with the previous program #4 example as shown in section 4.5.1. From any metering display press
PROGRAM, 4, ENTER, EDIT, EDIT,
and you should see the following display:
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
1, ENTER, ENTER,
to select segment 1 of 1.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
.OOO2, ENTER,
to select a time .0002 seconds. This is the smallest time allowed for a time
based transient segment.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
80, ENTER, 80, ENTER, 80, ENTER,
to select 80 volts on all three phases.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1
Press
400, ENTER,
to select 400 Hz. output frequency.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1
Содержание UPC-1
Страница 1: ...UPC 3 UPC 1 PROGRAMMABLE CONTROLLERS FIRMWARE V4 25 AND LATER OPERATION MANUAL PACIFIC POWER SOURCE δ ...
Страница 2: ......
Страница 12: ...SECTION 1 GENERAL viii THIS PAGE INTENTIONALLY BLANK ...
Страница 17: ...SECTION 1 GENERAL 5 PACIFIC Figure 1 3a SCU UPC 3 Figure 1 3b UPC 3 Figure 1 3c UPC 1 ...
Страница 43: ...SECTION 4 OPERATION 31 PA CIFIC FIGURE 4 1 POWER SOURCE FRONT PANEL ...
Страница 74: ...SECTION 4 OPERATION 62 90 0 100 0 85 95 100 90 FIGURE 4 6 2 1 EDITED WAVEFORM ...
Страница 134: ...NOTES 122 NOTES ...
Страница 135: ...MODIFCATIONS 123 MODIFICATIONS MODIFICATIONS This section is reserved for any modifications that may be made ...