L-718-8-E
5.10 How to Change Cable Entry Orientation of Terminal Box (Separately Mounted Type)
Of the sensor unit, the orientation of the terminal box
’
s cable entry can be changed.
◆Procedure◆
①
Turn OFF flowmeter power.
②
Using a flat tool, slowly loosen the terminal box lid by turning
counterclockwise, and loosen further by hand.
(Be careful not to damage the finish.)
(
Fig.5.12
)
Inside, the structure of terminal block can be either A or B shown below. Follow procedure
③~⑤
according to the type of your terminal block.
Fig.5.12
Terminal block A
Terminal block B
③
Remove the isolation
cover.
(
Fig.5.13, Fig.5.14
)
④
Disconnect the cables
from the terminal block.
(
Fig.5.15, Fig.5.16
)
Fig.5.13
Fig.5.14
Fig.5.15
Fig.5.16
CAUTION: Rotating the terminal box without following the procedure will cause
sensor damage.
NOTE: Although the terminal box can be rotated by loosening four hex-socket head screws at the
adapter, the following procedure must be followed.
Isolation cover
Isolation cover