Flow Direction
Valve
Bypass Line
Valve
Valve
ULTRAmass MK
ULTRA
MK?
L-718-8-E
4.3.7 Bypass Loop
For maintenance and servicing sake, it is good practice to provide a bypass loop.
ULTRA
Hex Bolt
Gasket
Gasket
Flow Direction
Arrow
IN
OUT
4.4 Installation
4.4.1 Flanged Type
(1) In order to avoid strains on sensor flanges,
be sure to correct the flange face-to-face
dimension, tilt and off-center of the pipes
immediately before and after the sensor.
(2) Make sure of the flow direction by referring to
the flow direction plate.
(3) Align the sensor flange O.D. with the flange
O.D. of the pipeline, install gaskets, and
tighten hex bolts evenly.
CAUTION: Remove, before installation, protective seals on sensor flanges that are attached
before shipment from factory.
Fig.4.4 Bypass Loop
Fig.4.5 Installation of Flanged Type