MyMode Motion Switching
Genium - Bionic Prosthetic System
52 | Ottobock
MyMode Motion Switching
Genium - Bionic Prosthetic System
52 | Ottobock
Switching from MyModes into basic mode
INFORMATION
Switching back to basic mode by bouncing up and down three or more times cannot be deactivated. Thus, you can
access basic mode even if switching via bouncing is not enabled in X-Soft.
return to
basic mode*
* Returning to basic mode is also possible by connecting the battery charger
≥ 3 x
1s
1 x
1 x
1 x
1 x
bouncing recognised
for switching
1. 3x bouncing on the forefoot with an extended leg and contact with ground throughout (within 1 second)
2. Remove load from prosthesis for 1 second (no ground contact)
• Feedback from the prosthesis: 1 long vibration signal
and 1 long beep signal
.
3. Wait for 1 second
• Feedback from the prosthesis: 1 short beep signal and 1 short vibration signal .
Switching from basic mode to the first MyMode
(3x bounce/4x bounce activated)
INFORMATION
If only bouncing three times is activated in X-Soft, any number of bounces greater than or equal to 3 switches to
the first MyMode.
bouncing recognised
for switching
first MyMode
activated
Switching by bouncing must be activated in X-Soft
4x bouncing must be activated in X-Soft
3 x
1 x
1 x
1s
2 x
2 x
1. 3x bouncing on the forefoot with an extended leg and contact with ground throughout (within 1 second)
2. Remove load from prosthesis for 1 second (no ground contact)
• Feedback from the prosthesis: 1 long vibration signal
and 1 long beep signal
.
3. Wait for 1 second
• Feedback from the prosthesis: 2 short beep signals and 2 short vibration signals .
Switching from basic mode to the second MyMode
bouncing recognised
for switching
second
MyMode
activated
4 x*
1 x
1 x
1s
3 x
3 x
* Switching by bouncing must be activated in X-Soft
* 4x bouncing must be activated in X-Soft
1. 4x bouncing on the forefoot with an extended leg and contact with ground throughout (within 1 second)
2. Remove load from prosthesis for 1 second (no ground contact)
• Feedback from the prosthesis: 1 long vibration signal
and 1 long beep signal
.
3. Wait for 1 second
• Feedback from the prosthesis: 3 short beep signals and 3 short vibration signals .