68 | Ottobock
E-MAG Active
60
C B
A
To open the lock, loosen the two screws to the right and left of the cap.
Check whether dust or dirt has settled on the spring D or lock E. Clean with a brush if required
(Figure 61 and 62).
Are the plug contacts F (Figure 62) from the joint to the electronics clean?
61
D
E
62
F
I
G
H
Before reassembling the unit, reapply protective silicone grease to the plug contacts. A plastic
syringe with a short, cut-off cannula facilitates controlling the dosage.
INFORMATION
The magnet (G) and the electronics (H) must not be removed.
CAUTION
Risk of malfunction due to disassembly.
The actuating bolt (I) must not be loosened, since this can affect
the functionality of the joint.