Ottobock | 35
B400
Please see the instructions for use supplied with the battery
charger for further details on use and on the LED displays.
6.7.5 Charging the battery
WARNING
Discharge of explosive gases during battery charging
Burns due to explosion after a user error
►
Ensure sufficient ventilation in enclosed spaces.
►
Do not smoke or light fires.
►
Sparks must be avoided.
►
Do not cover the ventilation slits in the case.
►
Only use battery chargers which have been verified
and approved by the manufacturer for use with the
respective batteries (observe information on the battery
charger). Non-observance may result in explosion of
the battery, resulting in a health hazard.
►
Follow the battery manufacturer safety instructions.
WARNING
Sparks
Risk of injury due to fire or explosion
►
Switch the battery charger off and disconnect the
mains plug before you disconnect the battery.
NOTICE
Improper charging
Damage to the batteries due to user error
►
Avoid deep discharge of the battery. The manufacturer
does not assume any liability for damage due to deep
discharge.
►
Charge the battery as soon as possible if the last three
segments of the "Battery Capacity" LED indicator are
flashing.
►
Charge the batteries weekly if the power wheelchair is
not used for an extended period of time.
1) Turn the control unit on the power wheelchair off.
2) Plug the battery charger plug into the charging/pro
gramming receptacle on the power wheelchair control
panel (see Fig. 15).
3) Connect the battery charger to the mains plug and turn
it on.
→
The charging process starts automatically, and the
battery charge condition is indicated by the LED
indicator on the control panel and on the battery
charger.
4) Turn the battery charger off when the charging process
is complete.
5) Remove the battery charger mains plug.
6) Disconnect the battery charger plug from the control
panel.
Usage
Содержание B400b400
Страница 1: ...Instructions for Use User B400...
Страница 2: ......