50
Problem
Solution
Check that the Heel Switch is placed correctly under
your heel inside the Heel Sock (see section 7.5 on
page 35).
Check the connection between the Heel Switch and
the Control Unit (see section 7.5.2 on page 37).
Check that the Heel Switch works properly by press-
ing the switch with your fingers. If you cannot get a
stable response when you are walking, contact your
clinician or order a new Heel Switch.
If the Control Unit does not respond with light sig-
nals and/or audio signals when you activate the Heel
Switch:
1. Check the battery status of the Heel Switch (see sec-
tion 7.1.6 on page 23) and the Control Unit (see
section 7.1.5 on page 22).
2. Repeat pairing of the Heel Switch and Control Unit
(see section 7.1.4 on page 21).
3. Replace the Heel Switch.
If the Control Unit responds with light signals and/
or audio signals when you activate the Heel Switch:
Check that the Antenna plug is properly connected
to the Control Unit.
You do not get a suf-
ficient foot lift when
you are walking
Make sure that the stimulation intensity is high
enough (see section 7.1.2 on page 18). If it is
already at the maximum, your need for stimulation
may have changed since your last visit at the clinic.
Contact your clinician for reprogramming of your
ActiGait®.
Your foot is lifted too
much when you are
walking
Make sure that the stimulation intensity is low
enough (see section 7.1.2 on page 18). If it is
already at the minimum, your need for stimulation
may have changed since your last visit at the clinic.
Contact your clinician for reprogramming of your
ActiGait®.