46
Pairing the Control Unit and Heel Switch
All lights are switched on when you
press the pairing button for the Heel
Switch. When you press the Heel
Switch they are switched off
See section 7.1.4 on page 21.
8.2 How to Interpret the Audio Signals
The Control Unit has audio signals and a warning signal. A warning means that the
Control Unit will beep for 8 seconds whereas audio signals are single beeps. Warn-
ings can never be muted. A warning means that stimulation has stopped and you
must find out what the warning is about and respond accordingly.
In this section you can see the different audio signals and warnings. The Control
Unit only has one type of sound that can be short or long, single or continuous.
Therefore, a short or long beep can mean different things depending on the situa
-
tion. Most often the audio signals are used in combination with light signals making
it possible to interpret what the audio signal tells you.
Short beep
+
A short beep when you press the intensity down button. Only one stimulation
intensity light indicator is on:
The stimulation intensity is already at the minimum level. See section 7.1.2 on
page 18.
Содержание ActiGait
Страница 2: ...ActiGait Implantable Drop Foot Stimulator User Manual ...
Страница 67: ...66 ...
Страница 68: ...67 ...