23
The battery level of the Control Unit is normal and
you have power for at least one day of walking.
The battery level of the Control Unit is medium and
you have power for less than one day.
Red and the Control Unit beeps once every 30 sec-
onds: The battery level is low and you have power
for less than 2 hours. Charge the Control Unit as
soon as you can.
Red and flashing while other light indicators are
turned off and the Control Unit beeps for 8 seconds
with an interval of 30 seconds:
The Control Unit is out of power and stimulation
has stopped. Turn off the Control Unit until you can
charge it.
When the Control Unit battery is out of power, both this light indicator and the light
indicator for the Heel Switch battery will flash and the Control Unit will beep for 8
seconds, while all other light indicators are switched off. You can tell which of the
two batteries is low by checking which one of the two is red: the Control Unit battery
indicator or the Heel Switch battery indicator.
It is recommended that you charge the Control Unit every night to avoid running out
of power at an inconvenient time.
7.1.6 The Heel Switch Battery Indicator
The Heel Switch is powered by an internal battery. The Heel Switch battery cannot
be recharged or exchanged so when the battery is empty you must discard the whole
Heel Switch (see section 12.3 on page 53 about disposal). Remember to pair the
new Heel Switch with the Control Unit when you replace it (see section 7.1.4 on