Surges
according
to
IEC 61000-4-5
±1 kV voltage
(line-to-line voltage)
±1 kV voltage
(line-to-line voltage)
The quality of the supply
voltage should be equival
ent to a typical commercial
or hospital supply.
Voltage drops, short inter
ruptions and fluctuations of
the supply voltage accord
ing to IEC 61000-4-11
0% U
T
; ½ period at 0, 45,
90, 135, 180, 225, 270 and
315 degrees
0% U
T
; 1 period
70% U
T
; 25/30 periods,
single phase: at 0 degrees
0% U
T
; ½ period at 0, 45,
90, 135, 180, 225, 270 and
315 degrees
0% U
T
; 1 period
70% U
T
; 25/30 periods,
single phase: at 0 degrees
The quality of the supply
voltage should be equival
ent to a typical commercial
or hospital supply.
Magnetic field with supply
frequency
(50/60 Hz)
according
to
IEC 61000-4-8
30 A/m
30 A/m
Magnetic fields at the
mains frequency should be
equivalent to the typical
levels for commercial or
hospital environments.
NOTE: U
T
is the mains voltage before applying the test levels.
Directives and manufacturer’s declaration – electromagnetic interference immunity
The 743A250 Bionic Pro product is designed for operation in the electromagnetic environment specified below.
The customer or user of the product must ensure that it is operated in such an environment.
INTERFERENCE
IMMUNITY TESTS
IEC 60601 test level
Compliance level
ELECTROMAGNETIC
ENVIRONMENT – direct
ives
Conducted HF interfer
ence
according
to
IEC 61000-4-6
Radiated HF interference
according
to
IEC 61000-4-3
3 V
eff
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
6 V
eff
in ISM frequency
bands between 0.15 MHz
and 80 MHz 80% AM at
1 kHz
3 V
eff
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
6 V
eff
in ISM frequency
bands between 0.15 MHz
and 80 MHz 80% AM at
1 kHz
Portable and mobile wire
less equipment should not
be used at a lesser dis
tance from the product
including the lines than the
recommended safety dis
tance calculated using the
equation applicable for the
transmission frequency.
Recommended
safety
distance:
d = 1.2√P
d = 1.2√P for 80 MHz to
800 MHz
d = 2.4√P for 800 MHz to
2.7 GHz
where P is the nominal out
put of the transmitter in
watts (W) according to the
transmitter
manufacturer
and d is the recommended
safety distance in metres
(m). The field strength of
stationary radio transmit
ters at all frequencies
according to an on-site
investigation
a
should be
less than the compliance
level.
b
Interference is possible in
the vicinity of devices that
bear the following symbol.
NOTE 1: The higher value applies at 80 MHz and 800 MHz.
50
Technical data
743A250 Bionic Pro
Содержание 743A250
Страница 1: ...743A250 Bionic Pro Instructions for use 3...
Страница 2: ...2 743A250 Bionic Pro...
Страница 53: ...53 743A250 Bionic Pro...
Страница 54: ...54 743A250 Bionic Pro...
Страница 55: ...55 743A250 Bionic Pro...