10 Technical data
General technical data
Supply voltage
100–240 V AC/50–60 Hz
Weight (case with all product
components)
9 kg
Case dimensions
500 x 432 x 118 mm
Technical data for wireless technology
Wireless technology
WLAN
Frequency range
2.4 GHz + 5 GHz
Frequencies
2.412 GHz–2.472 GHz + 5.160 GHz–5.865 GHz
Modulation type
DSSS, CCK, OFDM, BPSK, QPSK, 16QAM, 64QAM
Wireless protocol
802.11b, 802.11g, 802.11n, 802.11a, 802.11ac
Output
< 20 mW
Ambient conditions
Operating temperature
10–40 °C
Ambient temperature for char
ging
0–45 °C
Relative humidity
30%–70%
Specifications for EMC testing for EMC qualified personnel
Directives and manufacturer’s declaration – electromagnetic interference
The 743A250 Bionic Pro product is designed for operation in an electromagnetic environment as specified below.
The customer or user of the 743A250 Bionic Pro product must ensure that the device is operated in such an
environment.
Interference measurements
Compliance
Electromagnetic environment –
directives
HF emissions according to CIS
PR11
Group 1
The 743A250 Bionic Pro product
uses HF energy exclusively for its
internal FUNCTIONING. Therefore
its HF emissions are very low and
interference with neighbouring elec
tronic devices is unlikely.
HF emissions according to CIS
PR11
Class B
Emission of harmonics according to
IEC 61000-3-2
Class A
Emission of voltage fluctuations/flick
er according to IEC 61000-3-3
Complies
The 743A250 Bionic Pro product is
intended for use in all facilities
including residential, and those con
nected directly to the PUBLIC SUP
PLY NETWORK, which also sup
plies buildings used as residences.
Directives and manufacturer’s declaration – electromagnetic interference immunity
The 743A250 Bionic Pro product is designed for operation in the electromagnetic environment specified below.
The customer or user of the product must ensure that it is operated in such an environment.
INTERFERENCE
IMMUNITY TESTS
IEC 60601 test level
Compliance level
ELECTROMAGNETIC
ENVIRONMENT – direct
ives
Electrostatic
discharge
(ESD)
according
to
IEC 61000-4-2
±8 kV contact discharge
±15 kV air discharge
±8 kV contact discharge
±15 kV air discharge
Floors should be wood,
concrete or ceramic tile.
The relative humidity must
be at least 30%.
Electrical
fast
transi
ents/bursts according to
IEC 61000-4-4
± 2 kV for power lines
± 2 kV for power lines
The quality of the supply
voltage should be equival
ent to a typical commercial
or hospital supply.