background image

6) 

In case of collision:

Check  the  positioning  of  the  product  on  the  prosthetic  knee  joint  and
correct.
If the collision cannot be remedied by changing the positioning, qualified
personnel must remedy the collision according to the section "Prepara­
tion  for  use"  (see  Page 12)  –  heading  "Fitting  the  product  to  the  pros­
thesis".

Handling during charging

1)  Starting  from  the  medial  side,  thread  the  charging  adapter 

  between

the prosthetic knee joint 

 and the product 

 (see fig. 4).

2)  Verify  that  the  charging  adapter 

  is  correctly  fitted  to  the  prosthetic

knee joint   (see fig. 5).

3)  Charge the prosthetic knee joint according to its instructions for use.

Removing the product from the prosthesis

1)  Pull the rubber closure   off the pad button   (see fig. 6).
2) 

NOTICE! Do not overstretch the product.

Bend the product apart (see fig. 7).

3) 

NOTICE! Work carefully so as to not damage the prosthetic joint.

Remove the product from the prosthetic knee joint.

6.2 Cleaning

NOTICE

Use of unsuitable cleaning agents or disinfectants

Impairment of functionality and damage due to incorrect cleaning agents or
disinfectants

Only clean the product with the approved cleaning agents.

Only disinfect the product with the approved disinfectants.

Observe the instructions for cleaning and care.

Inform the patient.

Approved cleaning agents

Washing-up liquid, neutral soap

Approved disinfectants

Hand sanitiser (fragrance-free, alcohol-free)

14  |  Ottobock

Содержание 4X889-1

Страница 1: ...4X889 1 Gebrauchsanweisung 4 Instructions for use 10...

Страница 2: ...1 2 3 4 2 Ottobock...

Страница 3: ...5 6 7 8 Ottobock 3...

Страница 4: ...ichen ist 1 2 Kombinationsm glichkeiten Nachfolgend sind Prothesenkomponenten aufgef hrt die sich besonders zur Kombination mit dem Produkt eignen Benennung Kennzeichen Prothesenkniegelenk 3B1 3B1 ST...

Страница 5: ...Prothesenkompo nenten ob sie auch untereinander kombiniert werden d rfen VORSICHT Mechanische Besch digung des Produkts Verletzungen durch Funktionsver nderung oder verlust Arbeiten Sie sorgf ltig mit...

Страница 6: ...kondensierender Umgebungsfeuch tigkeit Vermeiden Sie den Kontakt mit abrasiven Medien z B Sand Staub Setzen Sie das Produkt keinen Temperaturen unter 10 C und ber 60 C aus z B Sauna berm iger Sonnene...

Страница 7: ...des Produkts darf nicht mit proximalen Prothesenkomponenten kollidieren Bei Flexion darf der Prothesenschaft nicht mit dem Produkt kollidieren Die Prothese in Flexionsstellung und Extensionsstellung b...

Страница 8: ...zwischen dem Prothesenkniegelenk und dem Produkt einf deln siehe Abb 4 2 Den richtigen Sitz des Ladeadapters am Prothesenkniegelenk berpr fen siehe Abb 5 3 Das Prothesenkniegelenk gem seiner Gebrauch...

Страница 9: ...Sicherheitskontrollen durchf hren 7 1 Austausch des Knieschutzes 1 Den alten Knieschutz vom Produkt entfernen siehe Abb 8 2 Nur die Klebefl che mit einem entfettenden Reiniger von Kleber ckst n den r...

Страница 10: ...urements so its shape matches that of the unaffected leg 1 2 Combination Possibilities Prosthetic components that are especially well suited for combination with the product are listed below Designati...

Страница 11: ...ilities Based on the instructions for use of the prosthesis components verify that they may be combined with each other CAUTION Mechanical damage to the product Injuries due to changes in or loss of f...

Страница 12: ...ratures below 10 C 14 F or above 60 C 140 F e g sauna excessive sunlight drying on a radiator Inform the patient 4 Scope of Delivery The following single components and accessories are included in the...

Страница 13: ...ollision 3 In case of collision Shorten and deburr the product in the collision area Avoid sharp edges and damage to the fibre structure 6 Use 6 1 Handling Applying product to the prosthesis 1 NOTICE...

Страница 14: ...rding to its instructions for use Removing the product from the prosthesis 1 Pull the rubber closure off the pad button see fig 6 2 NOTICE Do not overstretch the product Bend the product apart see fig...

Страница 15: ...ng cleaner 3 Fit the new knee pad and glue it using contact adhesive 4 Allow the adhesive to harden 8 Disposal This product may not be disposed of with regular domestic waste in all juris dictions Dis...

Страница 16: ...m in accordance with ISO 13485 Ottobock 647G979 DE EN 01 1408 Template Version 2014 05 28 FM477 SB_105x148 Otto Bock HealthCare GmbH Max N der Stra e 15 37115 Duderstadt Germany T 49 5527 848 0 F 49 5...

Отзывы: