4.11 Notes on the safety modes
CAUTION
Safety mode cannot be activated due to malfunction caused by water penetration or mechanical damage
Falling due to unexpected product behaviour because of changed damping behaviour.
►
Using the product when it is defective is prohibited.
►
The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety mode cannot be deactivated
Falling due to unexpected product behaviour because of changed damping behaviour.
►
If safety mode cannot be deactivated by recharging the battery, a permanent error has occurred.
►
Using the product when it is defective is prohibited.
►
The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected product behaviour because of changed damping behaviour.
►
The warnings/error signals (see Page 107) have to be observed.
►
After the safety signal has been emitted, further use of the product is prohibited.
►
The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
►
The warnings/error signals (see Page 107) have to be observed.
►
Particular caution is necessary when using a bicycle without a freewheel (with a fixed gear).
4.12 Information on the use of a mobile device with the cockpit app
CAUTION
Improper use of the mobile device
Falling due to altered damping behaviour as a result of unexpected switching into a MyMode.
►
Use the instructions for use (user) to instruct the patient on the proper handling of the mobile device with the
Cockpit app.
CAUTION
Independently applied changes or modifications made to the mobile device
Falling due to altered damping behaviour as a result of unexpected switching to a MyMode.
►
Do not make any independent changes to the hardware of the mobile device on which the app is installed.
►
Do not make any independent changes to the software/firmware of the mobile device that are not included in
the update function of the software/firmware.
CAUTION
Improper mode switching with the mobile device
Falling due to unexpected product behaviour because of changed damping behaviour.
►
Ensure that the patient stands securely during all switching processes.
►
Inform the patient that the changed damping characteristics have to be verified after switching, and feedback
from the acoustic signal emitter and the mobile device display must be observed.
►
Switching back to basic mode is mandatory once the activities in the MyMode have been completed.
75
C-Brace joint unit 17KO1=*
Safety