CAUTION
Independent manipulations on the joint unit and orthosis components
Falling due to the breakage of load-bearing components or malfunction of the orthosis.
►
Manipulations on the joint unit and orthosis components other than the tasks described in these instructions
for use are not permitted.
►
The battery may only be handled by Ottobock authorised, qualified personnel (no replacement by the user).
►
The joint unit and orthosis components may only be opened and repaired by authorised, qualified Ottobock
personnel.
CAUTION
Use of the product when battery charge level is too low
Falling due to unexpected behaviour of the product because of changed damping behaviour.
►
Check the current charge level before use and charge the product if required.
►
Note that the operating time of the product may be reduced at low ambient temperatures or due to ageing of
the battery.
CAUTION
Mechanical stress during transport
>
Falling due to unexpected product behaviour as a result of a malfunction.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
►
Only use the transport packaging for transportation.
CAUTION
Signs of wear and tear on the product components
Falling due to damage or malfunction of the product.
►
Regular service inspections (maintenance) are mandatory in the interest of patient safety and in order to main
tain operating reliability and protect the warranty.
CAUTION
Use of unapproved accessories
>
Falling due to product malfunction as a result of reduced interference resistance.
>
Interference of other electronic devices due to increased emissions.
►
Use the product only in combination with the accessories, signal converters and cables listed in the sections
"Scope of delivery" (see Page 76) and "Accessories" (see Page 76).
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
►
Clean the product with a damp cloth only (fresh water).
INFORMATION
Movement noises of the joint unit
When using orthotic joints, servomotor, hydraulic, pneumatic or brake load dependent control functions can
cause movement noises. This kind of noise is normal and unavoidable. It generally does not indicate any prob
lems. If movement noises increase noticeably during the lifecycle of the orthosis, the orthosis should be inspected
promptly by an authorised Ottobock Service Centre.
68
Safety