App. 3.3.9
Real time clock
Integrated real time clock for time-stamping of measurements. The clock is automatically synchronized with PC clock when
docked.
Accuracy:
Max. deviation 12 minutes/year
Backup:
Min. 7 days, when battery is removed from unit
App. 3.3.10
Data interfaces
Label printer:
RS232 interface to Label Printer
App. 3.3.11
Transport and storage environment
Temperature range:
-20 - +60°C (-4 - 140°F)
Humidity range:
10 - 90 % rel., non-condensing
Air pressure
500 hPa to 1040 hPa
App. 3.3.12
Operating environment
Mode of operation:
Continuous
Temperature range:
10 - 40°C (50 - 104°F)
Humidity range:
30 - 90 % rel., non-condensing
Air pressure
600 hPa* to 1040 hPa
*At locations where the normal air pressure is below 800 hPa (at altitudes of
more than 2000 meters), it is recommended to recalibrate the OAE probe.
App. 3.3.13
Warm-up time
< 20 seconds.
Note
•
Should be extended if MADSEN Alpha OAE has been stored in a cold environment.
App. 3.3.14
Disposal
MADSEN Alpha OAE can be disposed of as normal electronic waste, according to WEEE and local regulations.
Otometrics - MADSEN Alpha OAE
59
App. 3 Technical Specifications
Содержание MADSEN Alpha OAE
Страница 1: ...MADSEN Alpha OAE Screener User Manual Doc no 7 50 1370 US 01 Part no 7 50 13700 US 0 4 5 9 ...
Страница 6: ...6 Otometrics MADSEN Alpha OAE ...
Страница 10: ...1 Introduction 10 Otometrics MADSEN Alpha OAE ...
Страница 16: ...2 Unpacking and Installation 16 Otometrics MADSEN Alpha OAE ...
Страница 20: ...3 Getting Started with Alpha OAE 20 Otometrics MADSEN Alpha OAE ...
Страница 32: ...5 Testing with Alpha OAE 32 Otometrics MADSEN Alpha OAE ...
Страница 44: ...8 Handling Data in Alpha OAE 44 Otometrics MADSEN Alpha OAE ...
Страница 50: ...9 Maintenance and Cleaning 50 Otometrics MADSEN Alpha OAE ...
Страница 68: ......