Error message
Cause
Solution
Stimulus high
•
The probe has a poor fit in the
ear canal.
•
The probe is not in the ear
canal.
•
Try to refit the probe.
Stimulus low
•
The probe has a poor fit in the
ear canal.
•
The probe tip is blocked.
•
Try to refit the probe. If the problem per-
sists, clean the probe tip or replace it with
a spare.
Paper feed error
•
The labels are not adjusted cor-
rectly in the label printer or
there is a paper jam.
•
Re-adjust the labels and remove possible
blocking of paper feed.
56
Otometrics - MADSEN Alpha OAE
App. 2 Status and Error Messages