5.4
Options in the test result screens
Test result options
Print results
See
Test Other Ear
Starts test in the other ear.
Start
If you wish to repeat the test, press the
Start
button.
OK
Press to return to the
Home
screen.
5.5
Continuing testing on the other ear
1. Before you continue testing the other ear, check to make sure that the probe is clean. See also
.
2. If needed, reposition the patient so that you can test the other ear and insert the probe in the new
test ear.
3. Press the
Test Other Ear
button in the test results screen.
The test starts automatically.
5.6
Viewing a patients test record
A patient screened by Alpha OAE is either given a
PASS (Test passed)
or a
REFER (Test referred)
.
•
If the result is a
PASS (Test passed)
, no further action is required.
•
If the result is a
REFER (Test referred)
, it is recommended that you retest the patient. If the patient is given yet
another
REFER (Test referred)
, this patient should be referred for further testing.
5.6.1
Test View
You can view tests from the
Home
screen.
1. On the
Home
screen, press the
Test View
button.
This will take you to a list of tests.
30
Otometrics - MADSEN Alpha OAE
5 Testing with Alpha OAE
Содержание MADSEN Alpha OAE
Страница 1: ...MADSEN Alpha OAE Screener User Manual Doc no 7 50 1370 US 01 Part no 7 50 13700 US 0 4 5 9 ...
Страница 6: ...6 Otometrics MADSEN Alpha OAE ...
Страница 10: ...1 Introduction 10 Otometrics MADSEN Alpha OAE ...
Страница 16: ...2 Unpacking and Installation 16 Otometrics MADSEN Alpha OAE ...
Страница 20: ...3 Getting Started with Alpha OAE 20 Otometrics MADSEN Alpha OAE ...
Страница 32: ...5 Testing with Alpha OAE 32 Otometrics MADSEN Alpha OAE ...
Страница 44: ...8 Handling Data in Alpha OAE 44 Otometrics MADSEN Alpha OAE ...
Страница 50: ...9 Maintenance and Cleaning 50 Otometrics MADSEN Alpha OAE ...
Страница 68: ......