4
Preparing for testing
4.1
Preparing Alpha OAE
•
Follow the instructions in
Getting Started with Alpha OAE
•
See
on how to proceed with the specific tests.
4.2
Preparing the equipment
Every day before you start testing on patients, you should test the probe to make sure that it functions correctly. See
.
Caution
•
Insert only disinfected probe tips in the test cavity. See
Cleaning and disinfecting the test cavity
.
4.2.1
Preparing the probe for testing
Inspect the probe for deterioration (color changes, surface changes) of the reusable probe parts before every usage. If
deterioration occurs, contact your distributor.
4.2.2
Connecting the probe
1. Place Alpha OAE face down.
2. Align the double arrow on the probe plug with the arrow in the grey bordering around the socket of Alpha OAE.
3. Gently insert the probe plug into the probe socket on top of Alpha OAE. The plug does not require force to be inser-
ted properly.
Disconnecting the probe
Caution
•
Do not pull the plug by the cable when you disconnect the probe.
Otometrics - MADSEN Alpha OAE
21
4 Preparing for testing
Содержание MADSEN Alpha OAE
Страница 1: ...MADSEN Alpha OAE Screener User Manual Doc no 7 50 1370 US 01 Part no 7 50 13700 US 0 4 5 9 ...
Страница 6: ...6 Otometrics MADSEN Alpha OAE ...
Страница 10: ...1 Introduction 10 Otometrics MADSEN Alpha OAE ...
Страница 16: ...2 Unpacking and Installation 16 Otometrics MADSEN Alpha OAE ...
Страница 20: ...3 Getting Started with Alpha OAE 20 Otometrics MADSEN Alpha OAE ...
Страница 32: ...5 Testing with Alpha OAE 32 Otometrics MADSEN Alpha OAE ...
Страница 44: ...8 Handling Data in Alpha OAE 44 Otometrics MADSEN Alpha OAE ...
Страница 50: ...9 Maintenance and Cleaning 50 Otometrics MADSEN Alpha OAE ...
Страница 68: ......