3.4
Connect Bone Conduction Transducer
Note
•
This step applies only to tests that require bone conduction stimulus.
A. Connect bone conductor to the Chartr EP 200 preamplifier.
Connect Bone Oscillator to Preamplifier
B. Place bone oscillator on patient. Place left or right masking insert in ear (optional).
Bone conduction ABR provides a differential diagnosis of the type of hearing loss (sensorineural, conductive, or mixed).
The system default is a 2-channel protocol to be used without masking. The contralateral response will not have a Wave I
present and a slightly longer latency for Wave V. Wave I is not always present in the ipsilateral response.
Listening Check
Perform a listening check to verify that the output of the bone oscillator is accurate. See
and
lish Normal Hearing Level Thresholds
.
30
Otometrics - ICS Chartr EP 200
3 Prepare Patient for Testing
Содержание ICS Chartr EP 200
Страница 1: ...ICS Chartr EP 200 Installation and Startup Guide Doc No 7 50 0570 EN 10 Part No 7 50 05700 EN ...
Страница 4: ...Otometrics ICS Chartr EP 200 4 ...
Страница 16: ...16 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 17: ...Otometrics ICS Chartr EP 200 17 3 Prepare Patient for Testing ...
Страница 20: ...20 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 21: ...Otometrics ICS Chartr EP 200 21 3 Prepare Patient for Testing ...
Страница 23: ...Otometrics ICS Chartr EP 200 23 3 Prepare Patient for Testing ...
Страница 24: ...24 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 25: ...Otometrics ICS Chartr EP 200 25 3 Prepare Patient for Testing ...
Страница 26: ...26 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 27: ...Otometrics ICS Chartr EP 200 27 3 Prepare Patient for Testing ...