3.2.4
Tips for Tiptrodes
The Tiptrode consists of two components: the eartip/electrode and the connector/cable assembly. The electrode is the
foam eartip wrapped with gold foil. The connector/cable has an insulated alligator clip for attaching the eartip/electrode to
the tubing for transmitting auditory stimuli and an insulated wire to conduct the electrode signal.
Attach the eartip/electrode to the connector. Squeeze the metal clip open and then slide the eartip/electrode on making
sure that the alligator clip is touching the gold foil.
Connect the tubing to the output of the insert earphone.
Remove the tubing that is currently on the insert earphone and replace it with the tubing on the Tiptrode. Do NOT
double the length of the tubing by attaching the two sets together.
Red tubing goes to the right insert phone and blue tubing to the left insert phone.
Attach the electrode leadwire to the patient cable. See montages for electrode placement options.
Note
•
See
for continuation of this electrode placement.
Otometrics - ICS Chartr EP 200
19
3 Prepare Patient for Testing
Содержание ICS Chartr EP 200
Страница 1: ...ICS Chartr EP 200 Installation and Startup Guide Doc No 7 50 0570 EN 10 Part No 7 50 05700 EN ...
Страница 4: ...Otometrics ICS Chartr EP 200 4 ...
Страница 16: ...16 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 17: ...Otometrics ICS Chartr EP 200 17 3 Prepare Patient for Testing ...
Страница 20: ...20 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 21: ...Otometrics ICS Chartr EP 200 21 3 Prepare Patient for Testing ...
Страница 23: ...Otometrics ICS Chartr EP 200 23 3 Prepare Patient for Testing ...
Страница 24: ...24 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 25: ...Otometrics ICS Chartr EP 200 25 3 Prepare Patient for Testing ...
Страница 26: ...26 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 27: ...Otometrics ICS Chartr EP 200 27 3 Prepare Patient for Testing ...