Due to patient variability, it is recommended that the first VEMP run be collected with the EMG values set wide open
(e.g., 0 for min and 999 for max). Position the patient to contract the SCM and collect one run. The average EMG will be
displayed in the calculation box (bottom left). Then set the min to 20 µV below the average and the max to 20 µV above
the average (e.g., if average is 85 µV, then the min = 65 µV and the max = 105 µV).
The min and max values should be the same when collecting data from the left and right SCM to ensure that the EMG
value is similar on both sides. The response average will only include sweeps when the EMG is in between the min and
max values or “Good” range. If the EMG is in the “Low” or “High” range, those sweeps will be rejected.
The monitor light will display the following based on the EMG level:
Low (blue) – EMG level is below the
Min
value
Good (green) – EMG level is between the
Min
value and the
Max
value
High (amber) – EMG level is above
Max
value
Connect VEMP Monitor to Preamp
Go to
3.4 Step 9 Connect Bone Conduction Transducer
Otometrics - ICS Chartr EP 200
29
3 Prepare Patient for Testing
Содержание ICS Chartr EP 200
Страница 1: ...ICS Chartr EP 200 Installation and Startup Guide Doc No 7 50 0570 EN 10 Part No 7 50 05700 EN ...
Страница 4: ...Otometrics ICS Chartr EP 200 4 ...
Страница 16: ...16 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 17: ...Otometrics ICS Chartr EP 200 17 3 Prepare Patient for Testing ...
Страница 20: ...20 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 21: ...Otometrics ICS Chartr EP 200 21 3 Prepare Patient for Testing ...
Страница 23: ...Otometrics ICS Chartr EP 200 23 3 Prepare Patient for Testing ...
Страница 24: ...24 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 25: ...Otometrics ICS Chartr EP 200 25 3 Prepare Patient for Testing ...
Страница 26: ...26 Otometrics ICS Chartr EP 200 3 Prepare Patient for Testing ...
Страница 27: ...Otometrics ICS Chartr EP 200 27 3 Prepare Patient for Testing ...