CHAPTER NINE
TEOAE test
88
Test Result
Description
TE SNR OK
The data collected has met the criteria set. The
optimum test setting will depend on your application
e.g. screening or clinical measurement
(see chapter
17
Test set-up
)
Note: The following results will only occur if a TE SNR OK is not obtained
No Valid OAE
The data collected has not met the set Stop criteria
and the test conditions were acceptable
Noisy
If the noisy data collected is three times greater than
the low noise data collected
Poor Probe Fit
If the final test stimulus level is outside the stimulus ok
range or if the final stimulus stability value is < 85%
Too Few Bands
If insufficient bands meet the set requirements
Stopped Too
If a user ends the test manually before the required
Soon
minimum amount of data has been collected
9.4
TEOAE
test results
The following table lists all possible test results and gives an explanation of
the circumstances under which each result would be shown.
TEOAE test