Guidance and manufacturer’s declaration – electromagnetic immunity
The Otoport OAE+ABR is intended for use in the electromagnetic environment specified below. The
customer or the user of the Otoport OAE+ABR should ensure that it is used in such an environment.
Immunity
IEC 60601 Compliance
Electromagnetic environment -
test
test level
level
guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Otoport
OAE+ABR including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
1 Vrms
1 Vrms
d
= 3.5√
P
IEC 61000-4-6
150 kHz
to 80 MHz
Radiated RF
1 V/m
1 V/m
d
= 3.5√
P
80 MHz to 800 MHz
IEC 61000-4-3
80 MHz
d
= 7√
P
800 MHz to 2.5 GHz
to 2.5 GHz
where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
The conducted and radiated RF immunity levels are less than the IEC 60601 test levels. This is because the
Otoport OAE+ABR depends for its operation on the measurement of very low level physiological signals.
Two physiological signals are measured: an electro-physiological (ABR) signal of about 0.01 FemtoWatts
(giving about 100nV RMS across 1K); and, an acoustic signal of about 0.1 FemtoWatt acoustic power
(giving about 700nV RMS across 5K input impedance). It is impractical to measure these signals in the
presence of RF energy at the IEC 60601 test levels.
At high levels of RF energy both the ABR and OAE tests will be paused by the artifact rejection functions of
the instrument - it will go into ‘reject mode’. If this occurs then either the source of RF energy should be re-
duced or the measurement should be taken further from the source. In practice this means either switching
off other equipment (for example, mobile phones and noisy wall plug power supplies) or moving to another
location.
Technical specifications
Otoport OAE+ABR
User Manual for Advance and Screener