CHAPTER THIRTY
Response detection
266
Standard (mean +2 Std) based SNR values for four test points
(e.g. 2-6kHz frequency range)
Minimum SNR required
Confidence
2 of 4 points
3 of 4 points
4 of 4 points
99%
2dB
0dB
0dB
99.9%
3dB
2dB
0dB
99.99%
6dB
4dB
0dB
Rapid (mean) based SNR values for six test points (e.g. 1-6kHz
or 1.5-8kHz frequency range)
Minimum SNR required
Confidence
2 of 6 3 of 6 4 of 6 5 of 6 6 of 6
99%
8dB 6dB 4dB 3dB 1dB
99.9%
10dB 8dB 6dB 4dB 2dB
99.99%
12dB 9dB 7dB 5dB 4dB
Standard (mean +2 Std) based SNR values for six test points
(e.g. 1-6kHz or 1.5-8kHz frequency range)
Minimum SNR required
Confidence
2 of 6 3 of 6 4 of 6 5 of 6 6 of 6
99%
2dB 0dB 0dB 0dB 0dB
99.9%
4dB 2dB 0dB 0dB 0dB
99.99%
6dB 3dB 1dB 0dB 0dB
For example, if a test protocol is selected that requires 2 bands out of 6
tested and a 99.9% confidence is desired with ‘rapid’ noise mode then a
SNR of at least 10dB must be specified.