258
29.3
ABR test
Stimulus level
5-60 dBHL
Stimulus rate
51.8-57.9Hz
Stimulus polarity
Alternating
Stimulus type
Click or Chirp (see below)
Stimulus repetition rate
51.8-57.9 Hz
Evaluation method
Fsp & Template Correlation
Sampling rate
25.6 kHz
Frame length
17.3-18.8ms
Amplifier gain
75dB
Amplifier CMRR
> 60dB at 100 Hz
Input impedance
1MΩ || 10 nF
Amplifier noise
20 nV / Hz @ 0.1-1 kHz
Input bandwidth
150-1000 kHz
Notch filter
None
Impedance sense current
0.2µA
Impedance sense signal
400 Hz square wave
Impedance test range
0 to 99 kΩ
Display
EEG-level, ABR detection probability
Electrodes
Disposable wet gel electrodes
(FDA 510(k) cleared)
Electrode Montage
Positive electrode
Forehead
Negative electrode
Ipsilateral mastoid or nape of neck
Common electrode
Back of shoulder, cheek, or contralateral
mastoid