Warnings
CAUTION: THE INTERNAL COMPONENTS CAN BE STATIC SENSITIVE. USE CAUTION WHEN
OPENING THE ENCLOSURE AND HANDLING INTERNAL COMPONENTS.
Do not open the enclosure when the device is energized.
Do not open the enclosure if an explosive gas atmosphere may be present.
The Otis Instruments WireFree OI-6900 Sensor Assembly is Class I Division I, Groups C&D
Certified. The assembly is able to maintain its certification at all times while in the field,
simply by using the non-intrusive calibration method which requires the use of an Otis
Instruments, Inc. distributed magnet. However, if the Moore lid is removed, for whatever
reason, the OI-6900 certification is not valid. To avoid invalidating the certification, once the
device is put in the field, always use the Otis Instruments, Inc. distributed magnet to ensure
non-intrusive calibration.
Strong magnetic fields may interfere with the non-intrusive magnetic switches. A strong
magnetic field may momentarily active a switch, or permanently disable the switch to the “on”
or “off” position.
Do not cover the hole in the calibration cup, as this will cause the calibration to be inaccurate.
5