Calibration cont...
3. Unscrew and remove the sensor rainguard from the sensor housing.
4. Replace the sensor rainguard with an Otis OI-410 Calibration Cup.
5. Apply a known calibration gas to the OI-410 Calibration Cup that is attached to the sensor housing.
6. The sensor's detection of gas will begin to climb in value as shown on the display screen.
7. Watch the display screen until the number displayed stops increasing (or after approximately 90 seconds).
21