Sensor Replacement cont...
3. With the sensor housing cap removed, the visible sensor will resemble the following illustration:
4. Using the thumb and forefinger, slide the sensor out of the device.
NOTE: Do not use any metal object to remove the sensor.
NOTE: Be careful to not remove the sensor board when removing the sensor.
5. Slide the new sensor into device, matching the sensor prongs to the
corresponding eyelets inside.
6. Screw the sensor housing cap back in place.
NOTE: Once the sensor has been changed the device must be re-calibrated (see page 18).