Setting Calibration cont...
8. Touch the magnet to
ADD
(increase) or
SUB
(decrease) to manipulate the reading on the display screen to match
that of the calibration gas.
EXAMPLE: If the calibration gas is 25 PPM and the number on the display screen is 22 PPM,
touch the magnet to
ADD
approximately three times.
9. The device is now calibrated.
10. Unscrew the OI-410 Calibration Cup.
11. Reattach (screw on) the sensor rainguard to the sensor housing.
12. Once Calibration is complete, proceed to the next step.