Battery Replacement
To ensure full-functionality, the battery should be replaced if the voltage is less than 3.0. To check the battery voltage,
refer to the Advanced Menu Section of this Operation Manual.
CAUTION: THE INTERNAL COMPONENTS CAN BE STATIC SENSITIVE. USE CAUTION WHEN
OPENING THE ENCLOSURE AND HANDLING INTERNAL COMPONENTS.
NOTE: Replacement of the battery should be done in a non-classified environment, where no explosive gas is
present.
The device uses an Otis size “D” Lithium 19AH battery with connector. New batteries should only be obtained from Otis
Instruments, Inc. or an affiliated distributor.
1. Power off the device by touching and holding an Otis Instruments, Inc. distributed magnet against the right side of
the device for four seconds to activate
SUB
.
2. Unscrew, remove, and set aside the explosion-proof
Moore lid.