Unit Info cont...
3. Once the Unit Info has been viewed, continue to the next step.
Background Gas Setting
NOTE: When using an O2 sensor, two background setting modes will be available—Low and High. When
in these modes, the Background Low will be called “BackgroundL”; the Background High will be called
“BackgroundH”. When the background reading is below the Background Low setting—or above
Background High setting—the unit will transmit every five seconds.
1. After the Unit Info has been viewed (see above), touch an Otis Instruments, Inc. distributed magnet to the top side
of the device to activate the
MENU
button.
2. The display screen should resemble the following illustration:
3. Touch the magnet to
ADD
(increase) or
SUB
(decrease) to adjust the Background Gas Setting.
NOTE: The minimum background level that can be set is 0.01. The maximum background level
that can be set is 10% of full scale.
4. Once the Background Gas Setting has been set, continue to the next step.
29